- Home
- Guides
- Dew point to ppm
- −100 °C
Conversion · frost point
−100 °C frost point in ppm
At atmospheric pressure, a frost point of −100 °C means the gas contains 0.014 ppmV of water vapour. Below are the same value in mass units and natural gas units, and what it becomes at line pressure.
At a glance
0.014 ppmV
Water content of a gas with a frost point of −100 °C, measured or expressed at atmospheric pressure (1.013 bar).
| Quantity | Value |
|---|---|
| Water content (by volume) | 0.014 ppmV |
| In air (by mass) | 0.0086 ppmW |
| In nitrogen (by mass) | 0.0089 ppmW |
| Mass concentration, 0 °C reference | 0.011 mg/Nm³ |
| Mass concentration, 15 °C reference | 0.011 mg/Sm³ |
| Natural gas units | 6.6e−04 lb/MMscf |
| Water vapour pressure | 0.0014 Pa |
Vapour pressure from the Sonntag (1990) equations over ice; no enhancement factor. Results are the same as the dew point calculator with default settings.
The same gas at line pressure
Compressing a gas raises its dew point, because the water vapour partial pressure rises with the total pressure. A gas with 0.014 ppmV of water has these pressure dew points:
| Line pressure | Pressure dew point | ISO 8573-1 water class (compressed air) |
|---|---|---|
| 4 bar(g) | −91.8 °C frost point | Class 1 |
| 7 bar(g) | −89.2 °C frost point | Class 1 |
| 10 bar(g) | −87.5 °C frost point | Class 1 |
| 20 bar(g) | −83.8 °C frost point | Class 1 |
| 50 bar(g) | −78.4 °C frost point | Class 1 |
The other way round: if −100 °C is the pressure dew point at 7 bar(g), for example from a compressed air dryer, the gas contains about 0.0017 ppmV of water and its atmospheric frost point is about −109.6 °C.
Frost point or dew point? Below 0 °C, water condenses as frost, and −100 °C here is a frost point. If an instrument reports the dew point over supercooled water instead, the same water content reads as about −104.7 °C dew point. Check which convention your analyzer and your specification use before comparing numbers.
Ultra-dry, trace-moisture range
This is the range of high-purity and electronic-grade gases, nitrogen and argon from air separation units, and gas leaving a well-regenerated molecular sieve dryer upstream of LNG liquefaction or other cryogenic units. At these levels the sample system, not the sensor, usually limits the measurement: use short electropolished stainless steel tubing, leak-tight fittings and a steady flow.
Below −90 °C frost point only metal-oxide sensors such as the YORK DPT series (to −110 °C) cover the range; chilled mirror references stop around −90 °C.
See also the full dew point to ppm table and calculator, the ISO 8573-1 air class checker and the moisture unit converter.
Need to measure it?
YORK metal-oxide dew point transmitters and portable meters measure from −110 °C frost point to +20 °C dew point; chilled mirror hygrometers down to −90 °C frost point serve as references. Tell us the gas, the expected dew point and the pressure, and our Dubai team will propose a model.