−100 °C frost point in ppm

At atmospheric pressure, a frost point of −100 °C means the gas contains 0.014 ppmV of water vapour. Below are the same value in mass units and natural gas units, and what it becomes at line pressure.

At a glance

0.014 ppmV

Water content of a gas with a frost point of −100 °C, measured or expressed at atmospheric pressure (1.013 bar).

QuantityValue
Water content (by volume)0.014 ppmV
In air (by mass)0.0086 ppmW
In nitrogen (by mass)0.0089 ppmW
Mass concentration, 0 °C reference0.011 mg/Nm³
Mass concentration, 15 °C reference0.011 mg/Sm³
Natural gas units6.6e−04 lb/MMscf
Water vapour pressure0.0014 Pa

Vapour pressure from the Sonntag (1990) equations over ice; no enhancement factor. Results are the same as the dew point calculator with default settings.

The same gas at line pressure

Compressing a gas raises its dew point, because the water vapour partial pressure rises with the total pressure. A gas with 0.014 ppmV of water has these pressure dew points:

Line pressurePressure dew pointISO 8573-1 water class (compressed air)
4 bar(g)−91.8 °C frost pointClass 1
7 bar(g)−89.2 °C frost pointClass 1
10 bar(g)−87.5 °C frost pointClass 1
20 bar(g)−83.8 °C frost pointClass 1
50 bar(g)−78.4 °C frost pointClass 1

The other way round: if −100 °C is the pressure dew point at 7 bar(g), for example from a compressed air dryer, the gas contains about 0.0017 ppmV of water and its atmospheric frost point is about −109.6 °C.

Frost point or dew point? Below 0 °C, water condenses as frost, and −100 °C here is a frost point. If an instrument reports the dew point over supercooled water instead, the same water content reads as about −104.7 °C dew point. Check which convention your analyzer and your specification use before comparing numbers.

Ultra-dry, trace-moisture range

This is the range of high-purity and electronic-grade gases, nitrogen and argon from air separation units, and gas leaving a well-regenerated molecular sieve dryer upstream of LNG liquefaction or other cryogenic units. At these levels the sample system, not the sensor, usually limits the measurement: use short electropolished stainless steel tubing, leak-tight fittings and a steady flow.

Below −90 °C frost point only metal-oxide sensors such as the YORK DPT series (to −110 °C) cover the range; chilled mirror references stop around −90 °C.

See also the full dew point to ppm table and calculator, the ISO 8573-1 air class checker and the moisture unit converter.