Free tool · Units
Moisture and oxygen unit converter
Convert moisture and oxygen readings between the units used in specifications, datasheets and analyzer displays. Moisture conversions take the carrier gas into account.
↧ Download offline version (single HTML file, works without internet)
- ppmV
- —
- ppmW (mg/kg)
- —
- mg/m³ (0 °C)
- —
- mg/Sm³ (15 °C)
- —
- lb/MMscf
- —
- Dew point at pressure
- —
- Dew point at 1 atm
- —
- Water vapour pressure
- —
- % O₂
- —
- ppm O₂
- —
- ppb O₂
- —
- mg/m³ (0 °C)
- —
- Partial pressure
- —
- Mole fraction
- —
Conversion factors
| From ppmV to | Multiply by | Basis |
|---|---|---|
| ppmW (mg/kg) | 18.015 / Mgas | Molar mass of the carrier gas (air 28.96, N₂ 28.01, H₂ 2.016, SF₆ 146.06) |
| mg/m³ | 0.8037 | 0 °C, 101.325 kPa (22.414 L/mol) |
| mg/Sm³ | 0.7619 | 15 °C, 101.325 kPa (23.645 L/mol) |
| lb/MMscf | 1 / 21.06 | 60 °F, 14.696 psia |
Dew point uses the water vapour pressure from Sonntag (1990), over ice below 0 °C, with ideal mixing: e = ppmV × 10⁻⁶ × P. The same gas has a different dew point at every pressure, so always state the pressure.
For oxygen, 1 % = 10,000 ppm = 10,000,000 ppb. Partial pressure is the mole fraction multiplied by the total pressure, which is what an electrochemical sensor responds to.
ppmW depends strongly on the carrier gas: 10 ppmV of water is about 6.2 mg/kg in air but about 89 mg/kg in hydrogen.
Frequently asked questions
How do I convert ppmV to ppmW for moisture?
Multiply by the molar mass of water (18.015) and divide by the molar mass of the carrier gas. In air, 1 ppmV of water is about 0.62 ppmW; in SF₆ it is about 0.12 ppmW.
How many mg/m³ is 1 ppm of water?
About 0.80 mg/m³ at 0 °C and 101.325 kPa, or about 0.76 mg/m³ at 15 °C. State the reference conditions, because they change the result.
How many ppm is 1 % oxygen?
1 % is 10,000 ppm. Trace oxygen analyzers usually report in ppm, and percent analyzers in % O₂.
More engineering tools
- Dew point calculatorConvert pressure dew point to atmospheric dew point, ppmV, mg/m³ and lb/MMscf, with a pipe condensation check.
- ISO 8573-1 air class checkerFind the ISO 8573-1 water class of your compressed air from a dew point reading at any pressure.
- Analyzer selectorAnswer five questions and get the YORK moisture, oxygen or CO₂ analyzer that fits your application.
- Nitrogen purge calculatorHow long, and how much nitrogen, to purge a vessel or pipeline down to a target O₂ level.
- Natural gas water content calculatorConvert water dew point at line pressure to lb/MMscf, mg/Sm³ and ppmV, and check against a pipeline specification.
- SF₆ moisture calculatorConvert SF₆ dew point to ppmV and mg/kg, find the dew point at filling pressure and check IEC limits.
- Hydrogen purity checker (ISO 14687)Check water, oxygen and other impurities against ISO 14687 Grade D for fuel-cell hydrogen.
- Sample line response timeTransport time and T95 response for a gas sample line, from tubing size, length, flow and pressure.
Comparing readings from different analyzers?
Unit mix-ups are one of the most common causes of disputes over gas quality. Our engineers can help you set up consistent measurement and reporting.