Guide
Dew point to ppm (ppmV) conversion table and calculator
At atmospheric pressure, a dew point of −40 °C equals about 127 ppmV of water, −60 °C about 10.7 ppmV and −70 °C about 2.6 ppmV. Use the calculator for any value and line pressure, or the table from −100 °C to +20 °C below.
Dew point to ppm calculator
Change the dew point to get ppmV, or type ppmV to get the dew point. Below 0 °C the value is a frost point. For line-pressure conversions and more units, use the dew point calculator.
Conversion table at 1 atm
ppmV is parts per million by volume of water vapour in the gas. Below 0 °C the reading is a frost point (vapour pressure over ice); from 0 °C upwards it is a dew point (over water). Saturation vapour pressure from Sonntag (1990).
| Dew / frost point | Phase | Water content (ppmV) |
|---|---|---|
| −100 °C | Frost point | 0.014 |
| −90 °C | Frost point | 0.095 |
| −80 °C | Frost point | 0.540 |
| −70 °C | Frost point | 2.58 |
| −60 °C | Frost point | 10.7 |
| −50 °C | Frost point | 38.8 |
| −40 °C | Frost point | 127 |
| −30 °C | Frost point | 375 |
| −20 °C | Frost point | 1,019 |
| −10 °C | Frost point | 2,565 |
| 0 °C | Dew point | 6,032 |
| +10 °C | Dew point | 12,121 |
| +20 °C | Dew point | 23,087 |
For other values, or to include line pressure, use the dew point calculator.
Every value in 5 °C steps
One page per value from −110 °C frost point to +20 °C dew point, each with ppmV, ppmW, mg/Nm³, lb/MMscf and the pressure dew point at 4 to 50 bar(g):
−110 °C · −105 °C · −100 °C · −95 °C · −90 °C · −85 °C · −80 °C · −75 °C · −70 °C · −65 °C · −60 °C · −55 °C · −50 °C · −45 °C · −40 °C · −35 °C · −30 °C · −25 °C · −20 °C · −15 °C · −10 °C · −5 °C · 0 °C · +5 °C · +10 °C · +15 °C · +20 °C
How to calculate it
Water content in ppmV equals the saturation vapour pressure at the dew point divided by the total pressure of the gas, multiplied by one million: ppmV = e(Td) ÷ P × 10⁶. Because P is in the denominator, the same gas has a higher dew point when it is under pressure.
Worked example: compressed air
Compressed air at 7 bar(g) with a pressure dew point of −40 °C contains about 16 ppmV of water. If that air is expanded to atmospheric pressure, its dew point falls to about −57 °C. When you compare readings, always state whether the dew point is at line pressure or at atmospheric pressure.
Pressure dew point chart (compressed air at 7 bar(g))
The same air has a higher dew point under pressure. At 7 bar(g), the ISO 8573-1 humidity classes convert as follows:
| Pressure dew point at 7 bar(g) | ISO 8573-1 humidity class | Dew point at 1 atm | ppmV |
|---|---|---|---|
| −70 °C | Class 1 | −83 °C | 0.3 |
| −40 °C | Class 2 | −57 °C | 16 |
| −20 °C | Class 3 | −40 °C | 129 |
| +3 °C | Class 4 | −21 °C | 946 |
For other pressures, use the ISO 8573-1 air class checker.
Related
- Natural gas processing: measurement points (animated)
- Chilled mirror vs metal oxide: which dew point sensor?
- Compressed air dew point (ISO 8573-1)
- Natural gas water dew point
- Dew point meters in the UAE
- DPT-600 Plus portable dew point meter
- DPT830 in-line dew point analyzer
- DM series chilled mirror hygrometers
- Low humidity and low dew point generators for calibration
References
- The beginner's guide to humidity measurement (Good Practice Guide 124)National Physical Laboratory (NPL), UK, 2013. Introduces the concept of humidity and the basics of making a reliable humidity measurement.
- ISO 18453:2004 Natural gas — Correlation between water content and water dew pointInternational Organization for Standardization, 2004. Gives a method to convert between water content and water dew point in natural gas; confirmed as current in 2024.
- ISO 6327:1981 Gas analysis — Determination of the water dew point of natural gas — Cooled surface condensation hygrometersInternational Organization for Standardization, 1981. The standard for measuring natural-gas water dew point with chilled-mirror (cooled-surface) hygrometers; confirmed as current in 2024.
- ISO 8573-1:2010 Compressed air — Part 1: Contaminants and purity classesInternational Organization for Standardization, 2010. The standard that defines compressed-air purity classes, including water expressed as pressure dew point.
External sources open on the publisher's website. YORK Instrument is not affiliated with them. All measurement references →
Frequently asked questions
How many ppmV is a dew point of −40 °C?
At 1 atm, a frost point of −40 °C corresponds to about 127 ppmV of water vapour.
How many ppmV is a dew point of −70 °C?
At 1 atm, a frost point of −70 °C corresponds to about 2.58 ppmV.
Why does the dew point change with pressure?
Water content in ppmV is the ratio of water vapour pressure to total pressure. At higher line pressure the same ppmV gives a higher dew point, so a pressure dew point is always higher than the dew point of the same gas expanded to atmospheric pressure.
Is it dew point or frost point below 0 °C?
Below 0 °C water vapour normally condenses as frost, so the reading is a frost point. This table uses the vapour pressure over ice below 0 °C and over water above it.
How do I convert dew point to ppm?
Find the saturation vapour pressure at the dew point (over ice below 0 °C), divide it by the total gas pressure and multiply by one million: ppmV = e(Td) ÷ P × 10⁶. At 1 atm, −40 °C gives about 127 ppmV.
Need to measure dew point?
Tell us the gas, pressure and dew point range, and we will recommend a YORK instrument.