−30 °C frost point in ppm

At atmospheric pressure, a frost point of −30 °C means the gas contains 375 ppmV of water vapour. Below are the same value in mass units and natural gas units, and what it becomes at line pressure.

At a glance

375 ppmV

Water content of a gas with a frost point of −30 °C, measured or expressed at atmospheric pressure (1.013 bar).

QuantityValue
Water content (by volume)375 ppmV
In air (by mass)233 ppmW
In nitrogen (by mass)241 ppmW
Mass concentration, 0 °C reference301 mg/Nm³
Mass concentration, 15 °C reference286 mg/Sm³
Natural gas units17.8 lb/MMscf
Water vapour pressure38.0 Pa

Vapour pressure from the Sonntag (1990) equations over ice; no enhancement factor. Results are the same as the dew point calculator with default settings.

The same gas at line pressure

Compressing a gas raises its dew point, because the water vapour partial pressure rises with the total pressure. A gas with 375 ppmV of water has these pressure dew points:

Line pressurePressure dew pointISO 8573-1 water class (compressed air)
4 bar(g)−13.6 °C frost pointClass 4
7 bar(g)−8.4 °C frost pointClass 4
10 bar(g)−4.7 °C frost pointClass 4
20 bar(g)+3.5 °C dew pointClass 5
50 bar(g)+16.8 °C dew point—

The other way round: if −30 °C is the pressure dew point at 7 bar(g), for example from a compressed air dryer, the gas contains about 47.4 ppmV of water and its atmospheric frost point is about −48.4 °C.

Frost point or dew point? Below 0 °C, water condenses as frost, and −30 °C here is a frost point. If an instrument reports the dew point over supercooled water instead, the same water content reads as about −33.1 °C dew point. Check which convention your analyzer and your specification use before comparing numbers.

Moderately dry gas

Typical of instrument and plant air in many facilities, membrane dryers, and natural gas after glycol dehydration when the dew point is expressed at atmospheric pressure. Metal-oxide transmitters and portable meters cover this range easily.

Metal-oxide transmitters and portable meters (YORK DPT series) cover this value for process and field use; a chilled mirror hygrometer is the usual reference for calibration in safe areas with non-flammable gases.

See also the full dew point to ppm table and calculator, the ISO 8573-1 air class checker and the moisture unit converter.