Conversion · frost point
−40 °C frost point in ppm
At atmospheric pressure, a frost point of −40 °C means the gas contains 127 ppmV of water vapour. Below are the same value in mass units and natural gas units, and what it becomes at line pressure.
At a glance
127 ppmV
Water content of a gas with a frost point of −40 °C, measured or expressed at atmospheric pressure (1.013 bar).
| Quantity | Value |
|---|---|
| Water content (by volume) | 127 ppmV |
| In air (by mass) | 78.8 ppmW |
| In nitrogen (by mass) | 81.5 ppmW |
| Mass concentration, 0 °C reference | 102 mg/Nm³ |
| Mass concentration, 15 °C reference | 96.5 mg/Sm³ |
| Natural gas units | 6.01 lb/MMscf |
| Water vapour pressure | 12.8 Pa |
Vapour pressure from the Sonntag (1990) equations over ice; no enhancement factor. Results are the same as the dew point calculator with default settings.
The same gas at line pressure
Compressing a gas raises its dew point, because the water vapour partial pressure rises with the total pressure. A gas with 127 ppmV of water has these pressure dew points:
| Line pressure | Pressure dew point | ISO 8573-1 water class (compressed air) |
|---|---|---|
| 4 bar(g) | −25.0 °C frost point | Class 3 |
| 7 bar(g) | −20.2 °C frost point | Class 3 |
| 10 bar(g) | −16.8 °C frost point | Class 4 |
| 20 bar(g) | −9.7 °C frost point | Class 4 |
| 50 bar(g) | +0.8 °C dew point | Class 4 |
The other way round: if −40 °C is the pressure dew point at 7 bar(g), for example from a compressed air dryer, the gas contains about 16.0 ppmV of water and its atmospheric frost point is about −57.0 °C.
Frost point or dew point? Below 0 °C, water condenses as frost, and −40 °C here is a frost point. If an instrument reports the dew point over supercooled water instead, the same water content reads as about −43.7 °C dew point. Check which convention your analyzer and your specification use before comparing numbers.
Dry gas: desiccant dryer range
Typical of heatless and heated desiccant dryers, instrument air in process plants, nitrogen from PSA generators and battery dry rooms. −40 °C pressure dew point is ISO 8573-1 class 2 for compressed air.
Metal-oxide transmitters and portable meters (YORK DPT series) cover this value for process and field use; a chilled mirror hygrometer is the usual reference for calibration in safe areas with non-flammable gases.
See also the full dew point to ppm table and calculator, the ISO 8573-1 air class checker and the moisture unit converter.
Need to measure it?
YORK metal-oxide dew point transmitters and portable meters measure from −110 °C frost point to +20 °C dew point; chilled mirror hygrometers down to −90 °C frost point serve as references. Tell us the gas, the expected dew point and the pressure, and our Dubai team will propose a model.