1. Home
  2. Guides
  3. Dew point to ppm
  4. −60 °C

Conversion · frost point

−60 °C frost point in ppm

At atmospheric pressure, a frost point of −60 °C means the gas contains 10.7 ppmV of water vapour. Below are the same value in mass units and natural gas units, and what it becomes at line pressure.

At a glance

10.7 ppmV

Water content of a gas with a frost point of −60 °C, measured or expressed at atmospheric pressure (1.013 bar).

QuantityValue
Water content (by volume)10.7 ppmV
In air (by mass)6.63 ppmW
In nitrogen (by mass)6.86 ppmW
Mass concentration, 0 °C reference8.57 mg/Nm³
Mass concentration, 15 °C reference8.13 mg/Sm³
Natural gas units0.506 lb/MMscf
Water vapour pressure1.08 Pa

Vapour pressure from the Sonntag (1990) equations over ice; no enhancement factor. Results are the same as the dew point calculator with default settings.

The same gas at line pressure

Compressing a gas raises its dew point, because the water vapour partial pressure rises with the total pressure. A gas with 10.7 ppmV of water has these pressure dew points:

Line pressurePressure dew pointISO 8573-1 water class (compressed air)
4 bar(g)−47.5 °C frost pointClass 2
7 bar(g)−43.5 °C frost pointClass 2
10 bar(g)−40.8 °C frost pointClass 2
20 bar(g)−35.0 °C frost pointClass 3
50 bar(g)−26.5 °C frost pointClass 3

The other way round: if −60 °C is the pressure dew point at 7 bar(g), for example from a compressed air dryer, the gas contains about 1.35 ppmV of water and its atmospheric frost point is about −74.3 °C.

Frost point or dew point? Below 0 °C, water condenses as frost, and −60 °C here is a frost point. If an instrument reports the dew point over supercooled water instead, the same water content reads as about −64.6 °C dew point. Check which convention your analyzer and your specification use before comparing numbers.

Very dry gas

Typical of molecular sieve dryer outlets, LNG and NGL pre-treatment, industrial gas purity checks and dry-gas generators for calibration. It is also the drier end of battery dry rooms. Response time becomes important: a long or cold sample line can delay the reading by an hour or more.

Metal-oxide transmitters and portable meters (YORK DPT series) cover this value for process and field use; a chilled mirror hygrometer is the usual reference for calibration in safe areas with non-flammable gases.

See also the full dew point to ppm table and calculator, the ISO 8573-1 air class checker and the moisture unit converter.

Need to measure it?

YORK metal-oxide dew point transmitters and portable meters measure from −110 °C frost point to +20 °C dew point; chilled mirror hygrometers down to −90 °C frost point serve as references. Tell us the gas, the expected dew point and the pressure, and our Dubai team will propose a model.