Conversion · frost point
−60 °C frost point in ppm
At atmospheric pressure, a frost point of −60 °C means the gas contains 10.7 ppmV of water vapour. Below are the same value in mass units and natural gas units, and what it becomes at line pressure.
At a glance
10.7 ppmV
Water content of a gas with a frost point of −60 °C, measured or expressed at atmospheric pressure (1.013 bar).
| Quantity | Value |
|---|---|
| Water content (by volume) | 10.7 ppmV |
| In air (by mass) | 6.63 ppmW |
| In nitrogen (by mass) | 6.86 ppmW |
| Mass concentration, 0 °C reference | 8.57 mg/Nm³ |
| Mass concentration, 15 °C reference | 8.13 mg/Sm³ |
| Natural gas units | 0.506 lb/MMscf |
| Water vapour pressure | 1.08 Pa |
Vapour pressure from the Sonntag (1990) equations over ice; no enhancement factor. Results are the same as the dew point calculator with default settings.
The same gas at line pressure
Compressing a gas raises its dew point, because the water vapour partial pressure rises with the total pressure. A gas with 10.7 ppmV of water has these pressure dew points:
| Line pressure | Pressure dew point | ISO 8573-1 water class (compressed air) |
|---|---|---|
| 4 bar(g) | −47.5 °C frost point | Class 2 |
| 7 bar(g) | −43.5 °C frost point | Class 2 |
| 10 bar(g) | −40.8 °C frost point | Class 2 |
| 20 bar(g) | −35.0 °C frost point | Class 3 |
| 50 bar(g) | −26.5 °C frost point | Class 3 |
The other way round: if −60 °C is the pressure dew point at 7 bar(g), for example from a compressed air dryer, the gas contains about 1.35 ppmV of water and its atmospheric frost point is about −74.3 °C.
Frost point or dew point? Below 0 °C, water condenses as frost, and −60 °C here is a frost point. If an instrument reports the dew point over supercooled water instead, the same water content reads as about −64.6 °C dew point. Check which convention your analyzer and your specification use before comparing numbers.
Very dry gas
Typical of molecular sieve dryer outlets, LNG and NGL pre-treatment, industrial gas purity checks and dry-gas generators for calibration. It is also the drier end of battery dry rooms. Response time becomes important: a long or cold sample line can delay the reading by an hour or more.
Metal-oxide transmitters and portable meters (YORK DPT series) cover this value for process and field use; a chilled mirror hygrometer is the usual reference for calibration in safe areas with non-flammable gases.
See also the full dew point to ppm table and calculator, the ISO 8573-1 air class checker and the moisture unit converter.