1. Home
  2. Guides
  3. Dew point to ppm
  4. −80 °C

Conversion · frost point

−80 °C frost point in ppm

At atmospheric pressure, a frost point of −80 °C means the gas contains 0.540 ppmV of water vapour. Below are the same value in mass units and natural gas units, and what it becomes at line pressure.

At a glance

0.540 ppmV

Water content of a gas with a frost point of −80 °C, measured or expressed at atmospheric pressure (1.013 bar).

QuantityValue
Water content (by volume)0.540 ppmV
In air (by mass)0.336 ppmW
In nitrogen (by mass)0.347 ppmW
Mass concentration, 0 °C reference0.434 mg/Nm³
Mass concentration, 15 °C reference0.411 mg/Sm³
Natural gas units0.026 lb/MMscf
Water vapour pressure0.055 Pa

Vapour pressure from the Sonntag (1990) equations over ice; no enhancement factor. Results are the same as the dew point calculator with default settings.

The same gas at line pressure

Compressing a gas raises its dew point, because the water vapour partial pressure rises with the total pressure. A gas with 0.540 ppmV of water has these pressure dew points:

Line pressurePressure dew pointISO 8573-1 water class (compressed air)
4 bar(g)−69.8 °C frost pointClass 2
7 bar(g)−66.6 °C frost pointClass 2
10 bar(g)−64.3 °C frost pointClass 2
20 bar(g)−59.6 °C frost pointClass 2
50 bar(g)−52.9 °C frost pointClass 2

The other way round: if −80 °C is the pressure dew point at 7 bar(g), for example from a compressed air dryer, the gas contains about 0.068 ppmV of water and its atmospheric frost point is about −91.8 °C.

Frost point or dew point? Below 0 °C, water condenses as frost, and −80 °C here is a frost point. If an instrument reports the dew point over supercooled water instead, the same water content reads as about −84.9 °C dew point. Check which convention your analyzer and your specification use before comparing numbers.

Ultra-dry, trace-moisture range

This is the range of high-purity and electronic-grade gases, nitrogen and argon from air separation units, and gas leaving a well-regenerated molecular sieve dryer upstream of LNG liquefaction or other cryogenic units. At these levels the sample system, not the sensor, usually limits the measurement: use short electropolished stainless steel tubing, leak-tight fittings and a steady flow.

Metal-oxide transmitters and portable meters (YORK DPT series) cover this value for process and field use; a chilled mirror hygrometer is the usual reference for calibration in safe areas with non-flammable gases.

See also the full dew point to ppm table and calculator, the ISO 8573-1 air class checker and the moisture unit converter.

Need to measure it?

YORK metal-oxide dew point transmitters and portable meters measure from −110 °C frost point to +20 °C dew point; chilled mirror hygrometers down to −90 °C frost point serve as references. Tell us the gas, the expected dew point and the pressure, and our Dubai team will propose a model.