Meteorology: where to check humidity across a weather station network

Follow the humidity data from coastal, desert and airport stations to the instrument workshop and back to an accredited reference. Each numbered point shows what is checked, the typical conditions, why it matters and the instrument that fits.

COASTAL STATIONINLAND DESERT STATIONAIRPORT OBSERVING SYSTEMINSTRUMENT WORKSHOPACCREDITED LABSENSORS FOR CHECKPERIODIC CALIBRATIONRUNWAYHUMIDITY CHAMBERDM9200REFERENCESENSORS FROM NETWORKISO/IEC 170251H₂O2T3H₂O4H₂O5H₂O6DP
YORK instrumentIntegrated system (supplied through YORK)

Tap a numbered point. Swipe the diagram sideways on a phone.

Request a quote for this point →
Auto tour
Project list 0 →

How the network checks humidity

A weather service runs many stations and few references. The usual approach is a chilled mirror hygrometer used as a transfer reference: it travels to the stations for field comparisons, works in the instrument workshop, and is itself calibrated against a higher standard.

  1. Field stations. Coastal, inland and airport stations measure air temperature and relative humidity inside a radiation screen, usually with a capacitive polymer sensor.
  2. Field comparison. A technician places a chilled mirror reference beside the station sensor and logs both over a stable period.
  3. Sensor exchange. Sensors are rotated out of the network on a schedule and replaced with checked units.
  4. Workshop comparison. In the instrument workshop the removed sensors are compared with the reference in a humidity chamber at several set points.
  5. Reference calibration. The chilled mirror reference itself is calibrated at a regular interval by an accredited laboratory.
  6. Records. The results go into the station records, so data users can see when each sensor was checked and what was found.

Check points at a glance

#PointChecksTypical conditionsRecommended solution
1Coastal station: radiation screenRelative humidity and dew point (field comparison)Coastal summer dew points reach about 26–30 °C. At 35 °C air and 28 °C dew point, a 0.1 °C dew point error equals about 0.4 %RHDM9200
2Coastal station: air temperatureAir temperature next to the humidity sensorAt coastal humidity a 0.2 °C air temperature error shifts relative humidity by about 0.7–0.9 %RHDM9200 (PRT probe)
3Inland desert stationLow relative humidity and low dew pointOn summer afternoons inland, relative humidity can fall to about 5–8 %RH at 45 °C, a dew point of roughly −3 to +3 °CDM9200
4Airport observing systemDew point for aviation weather reportsRoutine aviation reports (METAR) give temperature and dew point in whole degrees CelsiusDM9200
5Workshop: humidity chamberSensor comparison against the referenceSeveral set points across the network's range, for example low, medium and high humidity at a stable temperature. Reference accuracy ±0.1 °C dew pointDM9200
6Traceability of the referenceCalibration of the transfer referenceCalibrated at a regular interval by an accredited laboratory (ISO/IEC 17025)Accredited calibration laboratory

Values are typical figures from published guidance and calculations from vapour pressure formulas. Your service's procedures and instrument specifications always govern.

Getting reliable comparisons

  • Pick calm conditions. Early morning or a calm overcast period gives steadier readings than a gusty afternoon.
  • Compare like with like. Convert both readings to dew point or to relative humidity using the same air temperature.
  • Let the mirror settle. Record only when the chilled mirror reading is stable.
  • Mind the dust. Inspect and clean the mirror and the sensor filter as part of every check.

References

  1. Guide to Instruments and Methods of Observation (WMO-No. 8)World Meteorological Organization
  2. A guide to the measurement of humidityNational Physical Laboratory (NPL)
  3. Chilled-mirror hygrometerAmerican Meteorological Society, Glossary of Meteorology

Frequently asked questions

Why check weather station humidity sensors in the field?

Capacitive humidity sensors drift slowly in heat, dust and salt air, and the readings still look plausible. A side-by-side comparison with an independent reference is the way to find the drift.

Why use a chilled mirror as the reference?

It measures dew point directly from the temperature at which condensation forms on a mirror, so it does not depend on a sensor material that ages.

Does air temperature matter in a humidity check?

Yes. Relative humidity depends on both dew point and air temperature, so both must be measured at the same place and time.