The Peltier cooler pulls heat out of the mirror.
At the dew point, dew or frost forms and scatters the light.
The controller keeps the film thickness constant.
The PRT reads the mirror: that temperature is the dew point.
Simplified illustration, time sped up. ppmV shown at 1 atm, calculated from the mirror temperature with the Sonntag (1990) equations (over ice below 0 °C, over water above).
Watch: the principle in one minute
A one-minute explainer: the mirror is cooled until dew or frost forms, the film is held steady, and the mirror temperature is the dew point.
English narration, with English and Chinese captions.
The principle in one sentence
Gas with a given water content has one temperature at which its water vapour starts to condense: the dew point (or, below 0 °C, the frost point). A chilled-mirror hygrometer finds that temperature physically, by cooling a surface until condensation appears and holding it there, and then measures the surface temperature.
What happens inside the sensor
- Cool. A thermoelectric (Peltier) cooler pumps heat from a small polished mirror into a heat sink. Deep dew points need more cooling stages.
- Condense. When the mirror reaches the dew point of the sample gas, a thin film of dew or frost forms. The film scatters the LED light, so less light reaches the detector.
- Hold. The controller uses the detector signal to adjust the cooling, so that the film neither grows nor evaporates. At that balance the mirror sits at the dew point.
- Read. A platinum resistance thermometer (PRT) embedded under the mirror measures its temperature. That temperature is the reading, and it can be converted to ppmV or other moisture units.
Why it is used as a reference
The measurement comes down to a temperature reading at the moment water condenses, not to a property of a sensing material. Its accuracy therefore depends mainly on the PRT and its calibration, which is why chilled-mirror hygrometers are widely used as reference and transfer standards, and to check other dew point sensors.
Dew or frost below 0 °C
Below 0 °C the condensate should be frost, but water can stay supercooled for a while. A supercooled dew reads lower than the true frost point: about 2 °C lower at a frost point of −20 °C, and about 3 °C at −30 °C. Good practice is to make sure the film has turned to ice before taking the reading.
In practice
- Keep the mirror clean: oil, particles and salts change how the film forms. Inspect and clean it as the manufacturer recommends.
- Use clean stainless steel sample lines and a steady flow. At very low dew points the reading takes longer to settle, because the line and mirror have to dry down.
- State whether the dew point is at line pressure or at atmospheric pressure when you compare readings.
Related
References
- Chilled-mirror hygrometerAmerican Meteorological Society, Glossary of Meteorology. Defines the instrument: a mirror is cooled until condensation scatters the reflected light, and electronics hold it at the dew point; accurate, with a slower response.
- A guide to the measurement of humidity (Good Practice Guide 103)National Physical Laboratory (NPL), UK, 1996. The UK national measurement institute's good practice guide for people who use humidity instruments.
- Hygrometers: calibration serviceNIST Thermodynamic Metrology Group. NIST calibrates chilled-mirror hygrometers in dew/frost point from −90 °C to +85 °C against its Hybrid Humidity Generator.
- Calibration of Hygrometers with the Hybrid Humidity Generator (NIST SP 250-83r1)C. W. Meyer, T. Herman, W. W. Miller, NIST, 2021. Describes the US national humidity standard and its uncertainty budget; chilled-mirror hygrometers serve as transfer standards in its international comparisons.
- Bilateral key comparison SIM.T-K6.5 on humidity standards in the dew/frost-point temperature range from −30 °C to +20 °CC. W. Meyer (NIST), A. Solano (LACOMET), Metrologia 53, Tech. Suppl. 03005, 2016. Two national laboratories compared their dew/frost-point standards using a chilled-mirror hygrometer shipped between them as the transfer standard; results agreed within uncertainty.
External sources open on the publisher's website. YORK Instrument is not affiliated with them. All measurement references →
Frequently asked questions
What does a chilled-mirror hygrometer measure?
It measures the dew point or frost point of a gas directly: the temperature at which water vapour in the gas condenses on a cooled mirror. The reading can then be converted to ppmV or other moisture units.
Why is the chilled mirror called a reference method?
Because the reading is a temperature measured at the moment of condensation, its accuracy depends mainly on the platinum thermometer under the mirror and its calibration, not on a moisture-sensitive material that can drift.
What is the difference between dew point and frost point?
Above 0 °C water vapour condenses as liquid dew; below 0 °C it normally condenses as frost. For the same gas, the dew point over supercooled water is a little lower than the frost point.
How low can a chilled mirror measure?
It depends on how much cooling the sensor has. Single-stage sensors cover moderate dew points; multi-stage or auxiliary cooling reaches much lower frost points. See the individual YORK models for their ranges.
Need a chilled-mirror hygrometer?
Tell us the gas, pressure and dew point range, and we will recommend a YORK instrument.