1. Home
  2. Guides
  3. Dew point to ppm
  4. −10 °C

Conversion · frost point

−10 °C frost point in ppm

At atmospheric pressure, a frost point of −10 °C means the gas contains 2,565 ppmV of water vapour. Below are the same value in mass units and natural gas units, and what it becomes at line pressure.

At a glance

2,565 ppmV

Water content of a gas with a frost point of −10 °C, measured or expressed at atmospheric pressure (1.013 bar).

QuantityValue
Water content (by volume)2,565 ppmV
In air (by mass)1,596 ppmW
In nitrogen (by mass)1,650 ppmW
Mass concentration, 0 °C reference2,062 mg/Nm³
Mass concentration, 15 °C reference1,954 mg/Sm³
Natural gas units122 lb/MMscf
Water vapour pressure260 Pa

Vapour pressure from the Sonntag (1990) equations over ice; no enhancement factor. Results are the same as the dew point calculator with default settings.

The same gas at line pressure

Compressing a gas raises its dew point, because the water vapour partial pressure rises with the total pressure. A gas with 2,565 ppmV of water has these pressure dew points:

Line pressurePressure dew pointISO 8573-1 water class (compressed air)
4 bar(g)+10.7 °C dew point—
7 bar(g)+17.9 °C dew point—
10 bar(g)+23.1 °C dew point—
20 bar(g)+34.2 °C dew point—
50 bar(g)+51.2 °C dew point—

The other way round: if −10 °C is the pressure dew point at 7 bar(g), for example from a compressed air dryer, the gas contains about 324 ppmV of water and its atmospheric frost point is about −31.4 °C.

Frost point or dew point? Below 0 °C, water condenses as frost, and −10 °C here is a frost point. If an instrument reports the dew point over supercooled water instead, the same water content reads as about −11.2 °C dew point. Check which convention your analyzer and your specification use before comparing numbers.

Refrigerant dryer and near-freezing range

Refrigerant compressed air dryers typically deliver a pressure dew point of about +3 °C (ISO 8573-1 class 4). Below 0 °C the measured value may be a frost point or a dew point over supercooled water; check which one your instrument reports.

Metal-oxide transmitters and portable meters (YORK DPT series) cover this value for process and field use; a chilled mirror hygrometer is the usual reference for calibration in safe areas with non-flammable gases.

See also the full dew point to ppm table and calculator, the ISO 8573-1 air class checker and the moisture unit converter.

Need to measure it?

YORK metal-oxide dew point transmitters and portable meters measure from −110 °C frost point to +20 °C dew point; chilled mirror hygrometers down to −90 °C frost point serve as references. Tell us the gas, the expected dew point and the pressure, and our Dubai team will propose a model.