Conversion · frost point

−15 °C frost point in ppm

At atmospheric pressure, a frost point of −15 °C means the gas contains 1,631 ppmV of water vapour. Below are the same value in mass units and natural gas units, and what it becomes at line pressure.

At a glance

1,631 ppmV

Water content of a gas with a frost point of −15 °C, measured or expressed at atmospheric pressure (1.013 bar).

QuantityValue
Water content (by volume)1,631 ppmV
In air (by mass)1,015 ppmW
In nitrogen (by mass)1,049 ppmW
Mass concentration, 0 °C reference1,311 mg/Nm³
Mass concentration, 15 °C reference1,243 mg/Sm³
Natural gas units77.4 lb/MMscf
Water vapour pressure165 Pa

Vapour pressure from the Sonntag (1990) equations over ice; no enhancement factor. Results are the same as the dew point calculator with default settings.

The same gas at line pressure

Compressing a gas raises its dew point, because the water vapour partial pressure rises with the total pressure. A gas with 1,631 ppmV of water has these pressure dew points:

Line pressurePressure dew pointISO 8573-1 water class (compressed air)
4 bar(g)+4.1 °C dew pointClass 5
7 bar(g)+10.9 °C dew point—
10 bar(g)+15.8 °C dew point—
20 bar(g)+26.3 °C dew point—
50 bar(g)+42.3 °C dew point—

The other way round: if −15 °C is the pressure dew point at 7 bar(g), for example from a compressed air dryer, the gas contains about 206 ppmV of water and its atmospheric frost point is about −35.6 °C.

Frost point or dew point? Below 0 °C, water condenses as frost, and −15 °C here is a frost point. If an instrument reports the dew point over supercooled water instead, the same water content reads as about −16.8 °C dew point. Check which convention your analyzer and your specification use before comparing numbers.

Refrigerant dryer and near-freezing range

Refrigerant compressed air dryers typically deliver a pressure dew point of about +3 °C (ISO 8573-1 class 4). Below 0 °C the measured value may be a frost point or a dew point over supercooled water; check which one your instrument reports.

Metal-oxide transmitters and portable meters (YORK DPT series) cover this value for process and field use; a chilled mirror hygrometer is the usual reference for calibration in safe areas with non-flammable gases.

See also the full dew point to ppm table and calculator, the ISO 8573-1 air class checker and the moisture unit converter.