Conversion · dew point
+15 °C dew point in ppm
At atmospheric pressure, a dew point of +15 °C means the gas contains 16,834 ppmV of water vapour. Below are the same value in mass units and natural gas units, and what it becomes at line pressure.
At a glance
16,834 ppmV
Water content of a gas with a dew point of +15 °C, measured or expressed at atmospheric pressure (1.013 bar).
| Quantity | Value |
|---|---|
| Water content (by volume) | 16,834 ppmV |
| In air (by mass) | 10,472 ppmW |
| In nitrogen (by mass) | 10,826 ppmW |
| Mass concentration, 0 °C reference | 13,530 mg/Nm³ |
| Mass concentration, 15 °C reference | 12,826 mg/Sm³ |
| Natural gas units | 799 lb/MMscf |
| Water vapour pressure | 1,706 Pa |
Vapour pressure from the Sonntag (1990) equations over water; no enhancement factor. Results are the same as the dew point calculator with default settings.
The same gas at line pressure
Compressing a gas raises its dew point, because the water vapour partial pressure rises with the total pressure. A gas with 16,834 ppmV of water has these pressure dew points:
| Line pressure | Pressure dew point | ISO 8573-1 water class (compressed air) |
|---|---|---|
| 4 bar(g) | +42.5 °C dew point | — |
| 7 bar(g) | +51.8 °C dew point | — |
| 10 bar(g) | +58.4 °C dew point | — |
| 20 bar(g) | +72.9 °C dew point | — |
| 50 bar(g) | +95.4 °C dew point | — |
The other way round: if +15 °C is the pressure dew point at 7 bar(g), for example from a compressed air dryer, the gas contains about 2,129 ppmV of water and its atmospheric frost point is about −12.1 °C.
Humid gas
Above the usual compressed air dryer range. Water can condense wherever the gas meets a surface colder than its dew point, for example in pipework exposed to air conditioning or at night.
Metal-oxide transmitters and portable meters (YORK DPT series) cover this value for process and field use; a chilled mirror hygrometer is the usual reference for calibration in safe areas with non-flammable gases.
See also the full dew point to ppm table and calculator, the ISO 8573-1 air class checker and the moisture unit converter.