Process guide · animated

Climatic test chambers: where to check humidity and temperature

Follow the air and steam into a temperature and humidity chamber, then out to a walk-in stability room. Each numbered point shows what is checked, the typical condition, why it matters and the instrument that fits.

DRY AIR SUPPLYHUMIDIFIERWALK-IN STABILITY ROOMCOMPRESSORDESICCANT DRYERH₂OREFRIGERATION UNITTEMPERATURE & HUMIDITY TEST CHAMBERDM9200TEMPERATURE LOGGERICH1DP2RH3T4DP5DP
YORK instrumentIntegrated system (supplied through YORK)

Tap a numbered point. Swipe the diagram sideways on a phone.

Request a quote for this point →
Auto tour
Project list 0 →

How the process works

Temperature and humidity chambers are used for electronics, automotive and materials testing, and walk-in rooms hold pharmaceutical stability samples. Test results are only as good as the condition the chamber really delivers.

  1. Temperature control. Heaters and a refrigeration unit hold the workspace temperature.
  2. Humidification. A steam humidifier adds water vapour to raise the humidity.
  3. Dehumidification. Dry air from a compressor and desiccant dryer, or a cold coil, removes water vapour for low humidity set points.
  4. Control. The chamber controller regulates to its own temperature and humidity sensors.
  5. Calibration. A chilled mirror reference and a temperature logger check the condition at the test position and across the workspace.

Check points at a glance

#PointChecksTypical conditionRecommended solution
1Chamber workspace: centreDew point reference at the test positionCommon test conditions include 40 °C / 93 %RH (IEC 60068-2-78, dew point about 38.6 °C) and 85 °C / 85 %RH (dew point about 81 °C). DM9200 range −30 °C FP to +90 °C DP, accuracy ±0.1 °CDM9200
2Chamber control sensorRelative humidity reading of the chamberAt 40 °C / 75 %RH a 0.1 °C dew point difference equals about 0.4 %RHDM9200
3Workspace temperature uniformityAir temperature at several positionsMapped at the corners and the centre of the workspace, as the test standard or customer requiresMulti-channel temperature logger (thermocouples or PRTs)
4Dry air supplyDew point of the dehumidification airWell below the chamber's lowest dew point. At 25 °C / 10 %RH the chamber dew point is about −8.7 °CDPT810, DPT-600 Plus
5Walk-in stability roomRoom humidity at the product shelvesICH Q1A storage conditions: 25 °C / 60 %RH (dew point about 16.7 °C), 30 °C / 65 %RH and 40 °C / 75 %RHDM9200

Values are typical figures from published standards and calculations from vapour pressure formulas. Your test standard and the chamber specification always govern.

Getting reliable readings

  • Measure at the test position. The chamber sensor sits near the air outlet; the test items sit elsewhere.
  • Avoid condensation. At high dew points, keep cables and sample parts warmer than the dew point.
  • Record temperature too. Relative humidity depends on temperature; record both at the same point.
  • Allow time. Humidity settles more slowly than temperature; wait for a stable period at each set point.

References

  1. A guide to the measurement of humidityNational Physical Laboratory (NPL)
  2. ICH Q1A(R2): Stability testing of new drug substances and productsInternational Council for Harmonisation (ICH)
  3. IEC 60068-2-78: Damp heat, steady stateInternational Electrotechnical Commission (IEC)

Frequently asked questions

How is a humidity test chamber calibrated?

A dew point reference and a temperature sensor are placed at the test position, the chamber is run at several set points, and its own reading is compared with the reference at each one. Temperature uniformity is mapped across the workspace.

Why use a chilled mirror for chamber calibration?

It measures dew point directly and covers the high dew points of damp heat tests, so it gives an independent reference for the chamber's own sensor.

What conditions are used for pharmaceutical stability rooms?

ICH Q1A lists 25 °C / 60 %RH, 30 °C / 65 %RH and 40 °C / 75 %RH among its storage conditions.

Calibrating chambers or stability rooms?

Tell us the chambers, set points and the tolerance you need, and we will propose a reference and a check routine.