Climatic test chambers: where to check humidity and temperature
Follow the air and steam into a temperature and humidity chamber, then out to a walk-in stability room. Each numbered point shows what is checked, the typical condition, why it matters and the instrument that fits.
Tap a numbered point. Swipe the diagram sideways on a phone.
How the process works
Temperature and humidity chambers are used for electronics, automotive and materials testing, and walk-in rooms hold pharmaceutical stability samples. Test results are only as good as the condition the chamber really delivers.
- Temperature control. Heaters and a refrigeration unit hold the workspace temperature.
- Humidification. A steam humidifier adds water vapour to raise the humidity.
- Dehumidification. Dry air from a compressor and desiccant dryer, or a cold coil, removes water vapour for low humidity set points.
- Control. The chamber controller regulates to its own temperature and humidity sensors.
- Calibration. A chilled mirror reference and a temperature logger check the condition at the test position and across the workspace.
Check points at a glance
| # | Point | Checks | Typical condition | Recommended solution |
|---|---|---|---|---|
| 1 | Chamber workspace: centre | Dew point reference at the test position | Common test conditions include 40 °C / 93 %RH (IEC 60068-2-78, dew point about 38.6 °C) and 85 °C / 85 %RH (dew point about 81 °C). DM9200 range −30 °C FP to +90 °C DP, accuracy ±0.1 °C | DM9200 |
| 2 | Chamber control sensor | Relative humidity reading of the chamber | At 40 °C / 75 %RH a 0.1 °C dew point difference equals about 0.4 %RH | DM9200 |
| 3 | Workspace temperature uniformity | Air temperature at several positions | Mapped at the corners and the centre of the workspace, as the test standard or customer requires | Multi-channel temperature logger (thermocouples or PRTs) |
| 4 | Dry air supply | Dew point of the dehumidification air | Well below the chamber's lowest dew point. At 25 °C / 10 %RH the chamber dew point is about −8.7 °C | DPT810, DPT-600 Plus |
| 5 | Walk-in stability room | Room humidity at the product shelves | ICH Q1A storage conditions: 25 °C / 60 %RH (dew point about 16.7 °C), 30 °C / 65 %RH and 40 °C / 75 %RH | DM9200 |
Values are typical figures from published standards and calculations from vapour pressure formulas. Your test standard and the chamber specification always govern.
Getting reliable readings
- Measure at the test position. The chamber sensor sits near the air outlet; the test items sit elsewhere.
- Avoid condensation. At high dew points, keep cables and sample parts warmer than the dew point.
- Record temperature too. Relative humidity depends on temperature; record both at the same point.
- Allow time. Humidity settles more slowly than temperature; wait for a stable period at each set point.
References
- A guide to the measurement of humidityNational Physical Laboratory (NPL)
- ICH Q1A(R2): Stability testing of new drug substances and productsInternational Council for Harmonisation (ICH)
- IEC 60068-2-78: Damp heat, steady stateInternational Electrotechnical Commission (IEC)
Frequently asked questions
How is a humidity test chamber calibrated?
A dew point reference and a temperature sensor are placed at the test position, the chamber is run at several set points, and its own reading is compared with the reference at each one. Temperature uniformity is mapped across the workspace.
Why use a chilled mirror for chamber calibration?
It measures dew point directly and covers the high dew points of damp heat tests, so it gives an independent reference for the chamber's own sensor.
What conditions are used for pharmaceutical stability rooms?
ICH Q1A lists 25 °C / 60 %RH, 30 °C / 65 %RH and 40 °C / 75 %RH among its storage conditions.
Calibrating chambers or stability rooms?
Tell us the chambers, set points and the tolerance you need, and we will propose a reference and a check routine.