Glove boxes: where to measure trace oxygen and moisture

Follow the argon from the cylinders into the box, around the purifier loop and through the antechamber. Each numbered point shows what is measured, the typical value, why it matters and the instrument that fits.

GAS SUPPLYREGENERATION GASGAS PURIFIERVENT (REGEN)EXHAUSTREFILL LINEAr 5.0REGULATORH₂ / N₂GLOVE BOXANTECHAMBERVACUUM PUMPBLOWERCOPPER CATALYSTMOLECULAR SIEVE1O₂2H₂O3O₂4H₂O5O₂·H₂O6O₂
YORK instrumentIntegrated system (supplied through YORK)

Tap a numbered point. Swipe the diagram sideways on a phone.

Request a quote for this point →
Auto tour
Project list 0 →

How the process works

Glove boxes give an oxygen- and moisture-free atmosphere for lithium battery research, OLED and semiconductor materials, organometallic chemistry and special welding. The box only stays clean if the gas supply, the purifier and the transfer routine are all working.

  1. Gas supply. High-purity argon (or nitrogen) from cylinders or a bulk supply fills the box and the antechamber.
  2. Circulation. A blower draws the box atmosphere through the purifier and returns it to the box.
  3. Purification. A copper catalyst removes oxygen and a molecular sieve removes water.
  4. Regeneration. When the purifier is loaded it is isolated, heated and flushed with a hydrogen and nitrogen mixture that reduces the copper and drives off the water to vent.
  5. Transfers. Items go in and out through the antechamber, which is evacuated and refilled several times before the inner door opens.

Measurement points at a glance

#PointMeasuresTypical value / limitRecommended solution
1Argon supply after the regulatorTrace oxygen in the inert gasArgon 5.0 (99.999 %) lists about 2 ppm O₂ in a typical supplier datasheetPGA500, POA200
2Argon supply after the regulatorTrace moisture in the inert gasArgon 5.0 lists about 3 ppm H₂O, roughly −69 °C frost pointDPT830
3Box atmosphereTrace oxygenBelow 1 ppm is the common target for an inert-gas glove boxPGA500, POA200
4Box atmosphereTrace moistureBelow 1 ppm, about −76 °C frost pointDPT830
5Purifier outletTrace oxygen and moisture after the purifierShould read below the box atmosphere. Purifiers are typically regenerated every 3 to 6 months with 5 % H₂ in N₂ while heatedPGA500, DPT830
6Antechamber after refillResidual oxygen before opening the inner doorEvacuate and refill at least three times before opening the inner doorPOA200

Values are typical figures from published specifications and technical literature. Your equipment supplier's specification always governs.

Getting reliable readings

  • Measure the supply too. A clean box with a dirty supply wears out the purifier quickly.
  • Return the sample gas. Return the sample to the box and keep the flow low.
  • Use the outlet trend. The purifier outlet tells you when to regenerate before the box atmosphere rises.
  • Check after maintenance. Glove changes and window seal work are common leak sources; check O₂ and H₂O afterwards.

References

  1. Glovebox operation (JoVE Science Education)JoVE · Pennsylvania State University
  2. Argon 5.0 product datasheetLinde Gas
  3. Argon-filled gloveboxes: specificationWarsaw University of Technology

Frequently asked questions

What oxygen and moisture levels does a glove box need?

Below 1 ppm of both is the common target for an inert-gas glove box. Some applications set tighter limits.

Where should the analyzers be connected?

On the box atmosphere, with the sample returned to the box. A second point at the purifier outlet shows when regeneration is due.

Can one analyzer serve several glove boxes?

A portable trace oxygen analyzer can check several boxes in turn. For continuous control each box usually has its own in-line analyzers.

Setting up a glove box line?

Send us the gas, the number of boxes and the limits you need, and we will propose the analyzers and sample points.