O₂ · Oxygen
YORK POA200 portable trace oxygen analyzer
The POA200 is a battery-powered trace oxygen analyzer for spot checks from 0.01 ppm to 25% O₂. It is used to check the purity of nitrogen and argon, confirm that a purge or inerting is complete, and verify the atmosphere in glove boxes and inert gas lines in safe areas.

| Model | POA200 |
|---|---|
| Type | Portable trace oxygen analyzer |
| Range | 0.01 ppm to 25% O₂ |
| Principle | Electrochemical |
| Sample handling | Built-in metering valve, flow meter and 4-way selection valve |
| Data and power | USB-C data interface; rechargeable battery |
| Not suitable for | CO₂ background and acid gases (H₂S, HCl, SO₂); use the POA3500 in gas containing CO₂ |
| Hydrogen and hydrocarbons | Usable; sensor life is shorter, about 12 months |
| Areas | Safe areas |
| Lead time | 6 to 8 weeks |
| Warranty | 12 months from shipment |
| Regional contact | Dubai, UAE — covers the GCC |
Product catalog (PDF) · certificate copies are sent with quotations.
Where it is used
| Job | What you check |
|---|---|
| Nitrogen and argon purity | ppm O₂ in cylinders, bulk supply and generator outlets; see industrial gas purity |
| Purge and inerting | Confirm oxygen is low enough before work or start-up; see the nitrogen purge calculator |
| Glove boxes | Oxygen in the inert atmosphere |
| Inert gas supply to battery and electronics lines | Trace O₂ in nitrogen or argon at the point of use |
Getting a reliable trace reading
- Use short stainless steel sample lines and leak-tight fittings; air leaking in raises the reading.
- Purge the line and set a steady flow with the metering valve before reading.
- At low ppm levels allow time for the line and cell to settle.
- Do not use the POA200 in CO₂ or acid gases; they shorten cell life and affect the reading.
POA200 or another YORK oxygen analyzer?
| If you need… | Choose |
|---|---|
| Portable trace O₂ checks | POA200 |
| Portable percent O₂ checks | POA300 |
| O₂ in gas containing CO₂ (flue gas, CO₂ product) | POA3500 paramagnetic |
| Continuous trace O₂ monitoring | PGA500 in-line |
Ordering in the UAE and GCC
Send your enquiry to our Dubai team with:
- the gas and where it is measured;
- the expected oxygen range;
- whether the gas contains CO₂, acid gases (H₂S, HCl, SO₂), hydrogen or hydrocarbons;
- whether the measurement point is in a hazardous area;
- quantity.
We reply with a quotation for the POA200. Instruments are shipped from the factory with export documents, 6 to 8 weeks from order confirmation. Electrochemical cells are consumables; in hydrogen and hydrocarbons expect to replace the cell about every 12 months.
Questions about the POA200
What is the range of the POA200?
0.01 ppm to 25% O₂, electrochemical.
Can the POA200 measure oxygen in CO₂?
No. Use the POA3500 paramagnetic analyzer in gas containing CO₂.
Can it be used in hydrogen?
Yes, but the electrochemical cell has a shorter life in hydrogen and hydrocarbons, about 12 months.
Can the POA200 be used in a hazardous area?
Use it in safe areas. For hazardous areas ask our Dubai team; the PGA500-EX is the explosion-proof in-line option.
How do I get a quotation in the UAE?
Email dubai@yorkinstrument.com or WhatsApp +971 55 137 4866 with the gas, oxygen range, background components and quantity.