A lamp sends broadband infrared light through the sample cell.
CO₂ absorbs light at 4.26 µm: more CO₂, less light at the measuring detector.
A reference detector reads a wavelength CO₂ does not absorb, through the same optics.
The ratio cancels lamp and window changes and gives the CO₂ concentration.
Simplified illustration. The curve and numbers use a generic Beer–Lambert model for one cell length; real analyzers choose the cell length for their measuring range and are linearised and calibrated with reference gases.
Watch: the principle in one minute
A one-minute explainer: CO₂ absorbs infrared light at 4.26 µm, and a reference channel cancels lamp ageing and window dirt.
English narration, with English and Chinese captions.
The principle in one sentence
Carbon dioxide absorbs infrared light strongly in a narrow band around 4.26 µm, so the less infrared light at that wavelength gets through a sample, the more CO₂ the sample contains.
What happens inside the analyzer
- Emit. An infrared source sends broadband light through a sample cell that the gas flows through. "Non-dispersive" means the light is not split into a spectrum with a prism or grating: optical filters pick the wavelengths instead.
- Absorb. CO₂ molecules absorb part of the light at 4.26 µm. The absorption follows the Beer–Lambert law: it depends on how many CO₂ molecules the light meets along the cell, so the signal falls along a curve, not a straight line.
- Compare. A second detector, behind a filter at a wavelength CO₂ does not absorb, sees the same lamp, windows and cell. If the lamp ages or a window gets dirty, both signals fall together.
- Calculate. The analyzer divides the measuring signal by the reference signal, linearises the result and reports CO₂ in % or ppm.
Why the reference channel matters
Without a reference, anything that dims the light looks like more CO₂. In the model above, a slightly dirty window makes clean air read about 10 % CO₂. Because the reference channel sees the same loss, the ratio stays the same and the reading stays correct. This is why YORK CO₂ analyzers use a reference channel outside the CO₂ band.
Choosing the range
A long light path through the gas gives high sensitivity for low concentrations; a short path keeps the signal usable up to 100 % CO₂. The cell is designed for the measuring range, which is why CO₂ analyzers are offered in several ranges.
In practice
- Keep the sample clean and above its dew point: liquid water or oil on the windows reduces the light, and too much loss leaves the analyzer outside its working range.
- Absorption depends on the number of molecules in the cell, so changes in cell pressure and temperature affect the reading. Keep the sample flow and pressure steady, or use an analyzer that compensates.
- Check zero with a CO₂-free gas such as nitrogen and span with a certified CO₂ mixture, at the interval your quality system requires.
References
- Nondispersive infrared sensorWikipedia. General overview of NDIR: broadband infrared light, an optical filter at the gas's absorption band (4.26 µm for CO₂) and a reference path.
- Carbon Dioxide Sensor Module Based on NDIR TechnologyL. Zhou, Y. He, Q. Zhang, L. Zhang, Micromachines 12(7):845, 2021. Open-access paper on a dual-channel NDIR CO₂ module, with temperature and humidity compensation of the detector signals.
External sources open on the publisher's website. YORK Instrument is not affiliated with them. All measurement references →
Frequently asked questions
What does NDIR mean?
NDIR stands for non-dispersive infrared. The analyzer measures how much infrared light a gas absorbs at a characteristic wavelength, using optical filters rather than a prism or grating to select that wavelength.
At what wavelength does CO₂ absorb infrared light?
NDIR CO₂ analyzers measure the strong CO₂ absorption band at about 4.26 µm.
Why does an NDIR analyzer need a reference channel?
The reference detector reads a wavelength that CO₂ does not absorb, through the same source and windows. Dividing the measuring signal by the reference signal cancels lamp ageing and window contamination, which would otherwise read as a change in CO₂.
How is an NDIR CO₂ analyzer calibrated?
Zero is set with a CO₂-free gas, usually nitrogen, and span with a certified CO₂ reference mixture close to the top of the measuring range.