How an electrochemical oxygen analyzer works

Oxygen diffuses through a membrane and is reduced at the cathode, and the cell current tells you how much oxygen is in the gas. Change the O₂ level, then add a small leak to see why trace measurements need care.

SAMPLE IN TO VENT small air leak µA CATHODE O₂ + 2H₂O + 4e⁻ → 4OH⁻ MEMBRANE lets O₂ through ELECTROLYTE ANODE LOAD CURRENT — O₂ N₂ e⁻
Last 40 s— reading   - - O₂ in the gas
Analyzer reading
5.00ppm
O₂ actually in the gas5.00 ppm
Cell current (illustrative)—
STABLE
Sample gas
O₂ in the gas5 ppm
0.01 ppm25 % O₂
Try a fault
01 DIFFUSE

Oxygen in the sample diffuses through a thin membrane into the cell.

02 REDUCE

At the cathode each O₂ molecule takes up four electrons and is reduced.

03 FLOW

Electrons flow from the anode through the external circuit: the cell makes its own current.

04 READ

The current is proportional to O₂ partial pressure, and the analyzer shows it in ppm or %.

Simplified illustration, time sped up. The number of molecules drawn follows a logarithmic scale, not the real ratio. The reaction shown is for a typical alkaline galvanic cell; the current value is illustrative.

Watch: the principle in one minute

A one-minute explainer: oxygen diffuses through a membrane, is reduced at the cathode, and the cell current is proportional to oxygen partial pressure.

English narration, with English and Chinese captions.

The principle in one sentence

An electrochemical oxygen sensor is a small battery that runs on oxygen: the more oxygen reaches its cathode, the more current it makes, so the current measures the oxygen in the gas.

What happens inside the cell

  1. Diffuse. The sample flows past a thin membrane. Oxygen diffuses through it at a rate set by its partial pressure in the gas.
  2. Reduce. At the cathode, just behind the membrane, oxygen takes up electrons and is reduced. In a typical alkaline cell: O₂ + 2H₂O + 4e⁻ → 4OH⁻.
  3. Flow. At the anode the matching reaction releases electrons, which flow through the external circuit. The cell needs no external voltage to make this current.
  4. Read. Because the membrane limits how fast oxygen arrives, the current is proportional to the oxygen partial pressure. The analyzer converts it to ppm or %. YORK electrochemical analyzers cover from 0.01 ppm up to 25 % O₂ (trace) or 0 to 100 % O₂ (percent), depending on the model.

Why trace oxygen needs care

At a few ppm, the gas you measure can easily contain more oxygen from the sample system than from the process. In the model above, a small air leak adds about 6 ppm: invisible in air, but it turns a 1 ppm gas into a 7 ppm reading. Use clean stainless steel tubing and fittings, leak-check the sample line, and purge it well before you trust the first reading.

Pressure, cell life and gas composition

  • The cell responds to oxygen partial pressure, so a change in cell pressure changes the reading by the same proportion. Vent the cell to a steady pressure, usually atmosphere.
  • In a galvanic cell the anode is slowly consumed, so the cell has a finite life and is replaced periodically. Keeping a trace cell away from air when it is not measuring helps it last and recover quickly.
  • Some gases, such as high CO₂ or acid gases, shorten the life of standard cells. Tell us the full gas composition so we can choose the right cell.
  • For percent oxygen without a consumable cell, YORK also offers the paramagnetic POA3500. See how to choose an oxygen analyzer.

References

  1. Electro-galvanic oxygen sensorWikipedia. Explains the galvanic cell: oxygen is reduced at the cathode, a diffusion barrier makes the current proportional to oxygen partial pressure, and the anode is slowly consumed.
  2. Optical methods for sensing and imaging oxygen: materials, spectroscopies and applicationsX. Wang, O. S. Wolfbeis, Chemical Society Reviews 43:3666–3761 (open access), 2014. Open-access review of optical oxygen sensing (absorption, luminescence quenching) as an alternative to electrochemical cells.

External sources open on the publisher's website. YORK Instrument is not affiliated with them. All measurement references →

Frequently asked questions

How does an electrochemical oxygen sensor work?

Oxygen diffuses through a membrane to a cathode, where it is reduced. The matching reaction at the anode releases electrons, and the resulting current is proportional to the oxygen partial pressure in the gas.

Can an electrochemical analyzer measure ppm and percent oxygen?

Yes, with the right cell and electronics. YORK trace analyzers such as the PGA500 and POA200 cover 0.01 ppm to 25 % O₂, and the percent analyzers POA300 and PGA530 cover 0 to 100 % O₂.

Why does my trace oxygen reading stay high?

The most common causes are air leaks or permeable tubing in the sample line, and a line that has not been purged long enough. At ppm levels even a very small leak adds more oxygen than the process gas contains.

Does an electrochemical oxygen cell wear out?

Yes. In a galvanic cell the anode is slowly consumed, so the cell is replaced periodically. Its life depends on the oxygen level it sees and on the gas composition.