Free tool · Calibration
Span gas dilution calculator
Enter the cylinder concentration, the concentration you need and the total flow. The tool gives the span and diluent flows and the uncertainty of the mixture.
YORK Instrument Middle East engineering team · Published 27 Nov 2026
- Dilution ratio
- —
- Resulting concentration
- —
- Absolute uncertainty
- —
- Cylinder lasts
- —
Method
Resulting concentration C = (Cs·Qs + Cz·Qz) / (Qs + Qz), so Qs = Qt·(Ct − Cz) / (Cs − Cz).
Relative uncertainty: u(C)/C = √[(Ucyl/2)² + (Qz/Qt)²·(us² + uz²)], with each flow controller's accuracy treated as a rectangular distribution (÷√3). The result is multiplied by 2 for the expanded uncertainty.
Limits
- The diluent must be the same balance gas as in the cylinder (usually N₂).
- Adsorption on tubing walls is not included; it matters at ppb levels of oxygen and moisture, so purge the line well before reading.
- Cylinder duration is approximate: ideal gas, emptied to zero.
See the oxygen analyzer selection guide.
Frequently asked questions
How do I make 100 ppm from a 1000 ppm cylinder?
Mix 100 mL/min of span gas with 900 mL/min of zero nitrogen for a total of 1 L/min (a 10:1 dilution). With a 2 % cylinder and two flow controllers of ±1 % of reading, the expanded uncertainty is about 2.5 %.
Do oxygen analyzers need factory recalibration?
No. YORK oxygen analyzers are calibrated on site: the user buys span and zero gas and runs the calibration. Only dew point meters need factory recalibration.
What is a sensible maximum dilution ratio?
In one stage, keep it at or below about 100:1. Above that the span flow gets small and flow controller error and diluent impurities dominate. For very low targets use a lower-concentration cylinder or a two-stage dilution.
Which standard covers dilution with mass flow controllers?
ISO 6145-7 covers dynamic preparation of calibration gas mixtures with thermal mass-flow controllers; the uncertainty is evaluated following the GUM (JCGM 100:2008).
Calibrating oxygen analyzers on site?
We can help you choose span gases and flow ranges that suit your YORK analyzers.