Electrochemical vs paramagnetic
| Electrochemical | Paramagnetic | |
|---|---|---|
| How it measures | Oxygen diffuses to a cathode and is reduced; the cell current is proportional to O₂ partial pressure | Measures the magnetic effect of oxygen, which is strongly paramagnetic, in the sample gas |
| Best range | Trace ppm up to percent levels | Percent levels, 0–100% O₂ |
| Sensor life | The cell is consumed over time and is replaced periodically | Non-depleting measurement |
| YORK models | POA200 and PGA500 (0.01 ppm to 25%); POA300 and PGA530 (0–100%); PGA500-EX | POA3500 (0–100%) |
Four questions to answer first
- What range? Below about 1% O₂ and down to ppm, use an electrochemical trace analyzer. For 0–100%, either principle works.
- Portable or fixed? Portable for spot checks and purging (POA200, POA300, POA3500); in-line for continuous monitoring (PGA500, PGA530).
- What is in the gas? Tell us the background gas and any acid gases, solvents or moisture, so the sensor and sample system can be matched.
- Is the area hazardous? For hazardous areas, consider the PGA500-EX and check the certificate documents against your project requirements.
Frequently asked questions
Which oxygen analyzer is used for ppm oxygen in nitrogen?
An electrochemical trace analyzer. The YORK POA200 (portable) and PGA500 (in-line) cover 0.01 ppm to 25% O₂.
When should I choose a paramagnetic analyzer?
For percent-level oxygen where a non-depleting sensor is preferred, for example checking generator output or process streams at 0–100% O₂. The YORK POA3500 uses the paramagnetic principle.
What about combustion and flue gas?
Zirconia sensors are widely used for combustion control and flue gas. YORK oxygen analyzers are electrochemical and paramagnetic and are intended for gas purity, inerting and process gases.
Need help choosing?
Send us the gas, the oxygen range and the installation, and we will recommend a model.