  3. Aviation MRO

Application

Aviation maintenance: breathing oxygen and tyre nitrogen checks

Checking the purity and moisture of aviator's breathing oxygen, oxygen in tyre nitrogen and inerting gas, and shop and breathing air at maintenance bases, with YORK portable oxygen analyzers and dew point meters.

Why it matters

Airlines and maintenance providers in the Gulf service large fleets at their home bases. Breathing oxygen goes from the cylinder bank into servicing carts and then into crew and passenger systems; dry nitrogen fills tyres, landing gear struts and accumulators. Moisture in oxygen can freeze in regulators at altitude, and oxygen in tyre nitrogen supports combustion if a hot brake overheats a tyre.

These checks are made at many small points across a base, so portable analyzers matter as much as fixed ones.

Check points and limits

#PointWhat is checkedTypical limitYORK instrument
1Oxygen cylinder bankOxygen purityAviator's breathing oxygen, Type I gas (MIL-PRF-27210): at least 99.5 % O₂POA3500, POA300
2Oxygen cylinder bankMoisture in breathing oxygenMIL-PRF-27210 Type I: no more than 7 ppm water vapour, dew point −63.3 °CDPT-600 Plus, DPT830
3Oxygen servicing cartPurity and moisture at the point of deliveryThe same limits apply where the oxygen enters the aircraftPOA300, DPT-600 Plus
4Nitrogen for tyre inflationOxygen in tyre nitrogenFAA AC 20-97B: no more than 5 % oxygen by volume in the tyrePGA530, POA300
5Fuel tank inerting systemOxygen in nitrogen-enriched airInerting is based on keeping the ullage at or below about 12 % O₂POA300, PGA530
6Paint hangar breathing airOxygen in breathing airEN 12021: 21 ± 1 % O₂POA300, PGA530
7Shop compressed airPressure dew pointThe ISO 8573-1 class the workshop specifies, for example class 2 (≤ −40 °C)DPT810, DPT-600 Plus

Values are taken from published specifications and guidance. The aircraft maintenance manual and the operator's approved procedures always govern. CO and CO₂ in breathing air are covered by an integrated system supplied through YORK.

Taking reliable readings

  • Check at the point of use. Measure at the cart or panel that feeds the aircraft, not only at the bulk supply.
  • Use oxygen-clean sample lines on aviator's oxygen, and keep them capped between jobs.
  • Allow time at low dew points. A reading near −63 °C takes longer to settle than a compressed air check; purge the line until it is stable.
  • Check the oxygen analyzer in air (20.9 % O₂) before the job. The POA300 is for safe areas.

Questions we are asked

What moisture limit applies to aviator's breathing oxygen?

MIL-PRF-27210 Type I gaseous oxygen allows no more than 7 ppm of water vapour, a dew point of −63.3 °C, with at least 99.5 % oxygen.

How much oxygen is allowed in aircraft tyre nitrogen?

FAA AC 20-97B says to inflate with dry nitrogen so that the tyre contains no more than 5 % oxygen by volume.

Can one portable analyzer cover several checks at a base?

Yes. A portable oxygen analyzer such as the POA300 covers oxygen carts, tyre nitrogen and inerting checks in turn, and a portable dew point meter such as the DPT-600 Plus covers oxygen and shop air moisture.

Which YORK instrument measures moisture in breathing oxygen?

The DPT-600 Plus portable dew point meter reads from −110 °C frost point to +20 °C dew point, so it covers the −63.3 °C MIL-PRF-27210 limit. For continuous monitoring at a cylinder bank, use the DPT830 in-line analyzer.

Suggested YORK instruments

Equipping a maintenance base?

Tell us the oxygen and nitrogen points and the checks you run today, and we will propose portable and fixed analyzers.