Guide

Why low dew point readings go wrong

Most wrong dew point readings come from the sample system, not the sensor. These eight checks cover the usual causes and how to fix them.

1. Leaks let ambient moisture in

At −60 °C frost point the gas holds about 11 ppmV of water. Air at a +25 °C dew point, common in Gulf summers, holds about 31,285 ppmV, roughly 3,000 times more. A leak too small to hear can therefore raise the reading by many degrees. Leak-test every fitting, keep the number of joints low, and run the sample line at slight positive pressure so that any leak flows outwards.

2. The tubing adds or releases water

Nylon, polyurethane and rubber tubing let water vapour pass through the wall and hold water on their surfaces. They are fine for high dew points but not for very dry gas. For frost points below about −40 °C, use short runs of clean stainless steel tubing and stainless fittings.

3. The reading is taken before it has settled

When a sensor and sample line go from wet to dry, water held on their surfaces is released slowly. This takes minutes at −20 °C but can take hours at −70 °C and below. Watch the trend and record the value only when it has stopped falling. Keep the line short, keep the flow steady, and purge new lines before you rely on them.

4. Pressure dew point and atmospheric dew point are mixed up

Reducing the pressure lowers the dew point: −40 °C at 7 bar(g) is about −57 °C at atmospheric pressure. ISO 8573-1 classes refer to the pressure dew point. Always note the pressure at the sensor, and convert with the dew point calculator when you compare readings.

5. The sample line is colder than the dew point

With wet gas, such as natural gas, process gas or air before the dryer, water condenses in any part of the line that is colder than the dew point, and the instrument then reads low. Keep the whole line warmer than the dew point, with heat tracing where needed. In the Gulf, watch for lines that pass through air-conditioned shelters or cool down at night.

6. Contamination on the sensor or mirror

Compressor oil, glycol, methanol, particles and corrosive gases foul metal-oxide sensors and chilled mirrors. Fit filters or coalescers suited to the gas, and clean the mirror of a chilled mirror hygrometer as part of routine maintenance. Soluble contaminants on a mirror shift the temperature at which dew forms and so shift the reading.

7. Hydrocarbons are taken for water

In natural gas, heavier hydrocarbons can condense on a chilled mirror at temperatures close to the water dew point. The water dew point and the hydrocarbon dew point are different measurements and need to be reported separately.

8. The sensor has drifted

Metal-oxide sensors drift over time. Check them periodically against a reference, such as a chilled mirror hygrometer or a dew point generator, and recalibrate when the difference is outside your tolerance.

Quick diagnosis

SymptomMost likely causeFirst check
Reading is wetter than expected and slowly fallingNot settled yet, or tubing releasing waterWait for a stable trend; check tubing material
Reading is wetter than expected and steadyLeak, or permeable tubingLeak-test fittings; use stainless steel below −40 °C
Reading jumps when people or doors moveLeak near the sensorTighten and test fittings at the sensor
Reading is drier than expected with wet gasCondensation in a cold lineHeat the line above the dew point
Two instruments disagree by a fixed amountDifferent pressures, or driftCompare at the same pressure; check calibration
Slow response after an upsetContamination on the sensor or mirrorInspect filters; clean the mirror

Equipment that helps

  • Sampling and conditioning: DPSS-61XX gas sampling system and analyzer panels built to your sample point.
  • Reference checks: chilled mirror hygrometers DM9600P (portable) and DM9900 (laboratory).
  • Calibration: YKDC220 dew point calibration system and DC2000 low humidity generator.
  • Field checks: DPT-600 Plus portable dew point meter.

Frequently asked questions

Why is my dew point reading higher than expected?

The most common causes are a leak in the sample system, permeable plastic tubing, or a reading taken before the line and sensor have dried out. At low dew points even a very small leak adds a large amount of water.

What tubing should I use for dew point measurement?

For frost points below about −40 °C, use short runs of clean stainless steel tubing with stainless fittings. Plastic tubing lets water vapour through the wall and holds water on its surface.

How long does a dew point reading take to stabilise?

It depends on how dry the gas is and how wet the system was. Minutes are usually enough at −20 °C, but at −70 °C and below it can take hours. Record the value when the trend has stopped changing.