Guide

Chilled mirror vs metal oxide: which dew point sensor?

Use a chilled mirror when you need a reference measurement; use metal oxide for fast, compact, continuous monitoring down to −110 °C frost point. Many plants use both.

Watch: the principle in one minute

A one-minute explainer: water vapour enters a nano-porous oxide film and changes its capacitance, which the instrument converts to dew point.

English narration, with English and Chinese captions.

Side-by-side comparison

Chilled mirrorMetal oxide (nano-porous film)
How it measuresCools a mirror until dew or frost forms and measures the mirror temperatureMeasures the capacitance change as water adsorbs in a porous oxide film
What the reading isA direct measurement of dew or frost pointAn inferred dew point from a calibrated sensor signal
Typical roleReference instrument, calibration, verification of other sensorsContinuous in-line monitoring, transmitters, portable spot checks
YORK rangeDown to −90 °C frost point (DM9900); −30 °C to +90 °C (DM9200)−110 °C frost point to +20 °C dew point (DPT series)
Response at low dew pointsSlower: the mirror must reach equilibriumFaster, especially when moving from wet to dry
Care in useClean sample and mirror; periodic mirror cleaningPeriodic recalibration against a reference
Size and costLarger, higher costCompact, lower cost per measuring point

How to choose

  • Calibration laboratory or reference check: chilled mirror (DM9900, DM9800, DM9700).
  • Field verification: portable chilled mirror (DM9600P).
  • Fixed monitoring after a dryer or on a gas line: metal-oxide transmitter or analyzer (DPT810, DPT830).
  • Hazardous area: DPT990-EX.
  • Audits and commissioning: portable metal-oxide meter (DPT-600 Plus).

To check that your dew point sensors read correctly, YORK also makes humidity calibration systems such as the YKDC220 and the DC2000 low humidity generator.

For a comparison that includes laser analyzers, see chilled mirror vs TDLAS vs metal-oxide.

References

  1. Chilled-mirror hygrometerAmerican Meteorological Society, Glossary of Meteorology. Defines the instrument: a mirror is cooled until condensation scatters the reflected light, and electronics hold it at the dew point; accurate, with a slower response.
  2. Sensitivity Enhancement of a PPM Level Capacitive Moisture SensorL. Kumar, T. Islam, S. C. Mukhopadhyay, Electronics 6(2):41, 2017. Open-access paper on a porous aluminium-oxide capacitive sensor for 3 to 100 ppm moisture in process gases, the sensing principle behind metal-oxide dew point meters.
  3. A guide to the measurement of humidity (Good Practice Guide 103)National Physical Laboratory (NPL), UK, 1996. The UK national measurement institute's good practice guide for people who use humidity instruments.
  4. Calibration of Hygrometers with the Hybrid Humidity Generator (NIST SP 250-83r1)C. W. Meyer, T. Herman, W. W. Miller, NIST, 2021. Describes the US national humidity standard and its uncertainty budget; chilled-mirror hygrometers serve as transfer standards in its international comparisons.

External sources open on the publisher's website. YORK Instrument is not affiliated with them. All measurement references →

Frequently asked questions

Is a chilled mirror more accurate than a metal-oxide sensor?

A chilled mirror measures the dew point directly, so it is the usual choice as a reference and for calibration. A metal-oxide sensor gives an inferred value and relies on its calibration, but it is faster and better suited to many fixed measuring points.

Which sensor goes lower?

YORK metal-oxide analyzers cover down to −110 °C frost point. YORK chilled mirror hygrometers reach −90 °C frost point (DM9900).

Can I use both?

Yes. A common setup is metal-oxide transmitters at each process point and a chilled mirror hygrometer to verify them periodically.

Not sure which to choose?

Send us your gas, pressure and dew point range, and we will recommend the right principle and model.