- Mixing ratio
- —
- Water content
- —
- Vapour pressure
- —
- Dew point depression
- —
Method
The tool uses the Sonntag (1990) saturation vapour pressure equations over water and over ice, the same equations as every other YORK tool on this site. Relative humidity = e / ew(t) × 100, where e is the water vapour pressure and ew(t) is the saturation pressure over water at the air temperature. Absolute humidity = e × Mw / (R × T); mixing ratio = 622 × e / (P − e) g/kg.
Limits
- Valid for air temperatures from −45 to +60 °C and pressures from 50 to 200 kPa.
- Ideal-gas behaviour is assumed and the enhancement factor is ignored; at atmospheric pressure the error is below 0.5 %.
- For compressed gases use the dew point calculator; for natural gas use the natural gas water content calculator.
Where it is used
Climatic test chambers, battery dry rooms, date stores, and checking RH sensors against a chilled mirror reference. See measurement points in climatic test chambers.
Frequently asked questions
What is the dew point at 30 °C and 60 % relative humidity?
About 21.4 °C at atmospheric pressure. Any surface colder than that, such as a chilled-water pipe or an air-conditioned wall, will collect condensation.
What is the dew point on a Gulf summer day at 45 °C and 20 % RH?
About 16.8 °C, which is an absolute humidity of about 13.1 g/m³. The relative humidity looks low, but the air holds a lot of water, which is why cooled equipment sweats.
Why are dry gases measured as dew point and not relative humidity?
Relative humidity changes with temperature even when the amount of water stays the same, and most RH sensors lose accuracy below about 10 %RH. Dew point is an absolute measure of water content, which is why standards such as ISO 8573-1 for compressed air use it.
Below 0 °C, is relative humidity calculated over water or over ice?
By the WMO definition (WMO-No. 8), relative humidity is always referred to liquid water. Example: a dew point of −20 °C in air at 25 °C is about 4.0 %RH.
Need a reference for RH sensors?
We supply traceable chilled mirror hygrometers to use as a lab or field reference.