Semiconductor gases: where to measure moisture and oxygen

Follow bulk gas from the storage tank to the process tool, and clean dry air from the compressor to the fab. Each point shows what is measured and which analyzer fits.

ELECTROPOLISHED BULK LINEAIRCLEAN DRY AIRHOUSE N₂BULK LIQUIDN₂ / Ar / O₂H₂ TUBE TRAILERVAPORIZERINCOMINGQCPURIFIERVALVEMANIFOLD BOXFABPROCESSTOOLCDA COMPRESSORCDA DRYERN₂ GENERATOR1H₂O·O₂2H₂O·O₂3ppb4ppb5H₂O6O₂
YORK instrumentIntegrated system (supplied through YORK)

Tap a numbered point. Swipe the diagram sideways on a phone.

Request a quote for this point →
Auto tour
Project list 0 →

⤓ Download the 2-page PDF application note

How the process works

Fabs use bulk nitrogen, argon, oxygen and hydrogen at sub-ppb impurity levels. YORK instruments screen incoming gas, watch purifiers and check CDA and house nitrogen; ppb and ppt certification uses dedicated trace analyzers.

  1. Bulk supply. Liquid gases from on-site tanks are vaporized; hydrogen often arrives by tube trailer.
  2. Incoming QC. Each delivery is screened before it reaches the purifier.
  3. Purification. Point-of-entry purifiers take moisture, oxygen, CO₂ and hydrocarbons down to the sub-ppb level.
  4. Distribution. Electropolished stainless lines and valve manifold boxes carry the gas to each tool.
  5. Utilities. Clean dry air and house nitrogen feed pneumatic valves, purges and inerting across the fab.

Measurement points at a glance

#PointMeasuresTypical value / limitRecommended solution
1Incoming tube trailer / bulk tank checkMoisture and oxygen (screening)Before the purifier, moisture can exceed 10 ppmDPT830, POA200
2Purifier outlet (breakthrough watch)Moisture and oxygen (screening)Should be below the analyzer floor; any rise signals breakthroughDPT830, POA200
3Bulk distribution after the purifierH₂O, O₂, CO₂, CH₄Below 1 ppb; industry roadmap target below 100 pptIntegrated system: atmospheric-pressure ionization mass spectrometer (APIMS), cavity ring-down spectroscopy (CRDS) or a specialised trace GC
4Point of use at the toolH₂O, O₂, CO₂, CH₄Sub-ppb; consistency at the point of use matters more than the bulk numberIntegrated system: APIMS, CRDS or specialised trace GC
5Clean dry air (CDA) dryer outletPressure dew pointSet by the fab specificationDPT830, DPT810
6House nitrogen generator outletResidual oxygenSet by the fab specificationPOA200, PGA500

Values are typical figures from published specifications and technical literature. Your contract, licensor or plant specification always governs.

Practical notes

  • Know the floor. A trace oxygen analyzer reading to 0.01 ppm and a −110 °C frost point sensor are good screening tools, but they cannot certify ppt-level gas.
  • Moisture lingers. Moisture adsorbs on tube walls and is released slowly, so dry-down after maintenance takes time.
  • Materials matter. Polymer parts such as PTFE can release moisture for a long time; use electropolished stainless steel in the sample path.

References

  1. Bulk gas purity and APIMS measurementSolid State Technology / Semiconductor Digest

Frequently asked questions

Can YORK analyzers certify semiconductor-grade gas?

No. Sub-ppb and ppt certification needs APIMS, CRDS or specialised trace GC, which YORK can supply as an integrated system. YORK instruments are used for incoming screening, purifier breakthrough alarms, CDA and house nitrogen.

Why is moisture the hardest impurity in UHP gas?

Water adsorbs on the inside of tubes and components and is released slowly, so it persists long after the source is removed.

What does −110 °C frost point correspond to?

About 1.6 ppbv of water at atmospheric pressure (calculated).

Planning gas monitoring for a fab?

Tell us the gases, the points and the levels you need, and we will propose YORK instruments and integrated trace analyzers.