Application
Green hydrogen and ammonia: moisture and oxygen in hydrogen
Measuring oxygen and trace water in hydrogen from the electrolyser to the ammonia loop or refuelling dispenser, with YORK explosion-proof dew point and trace oxygen analyzers.
Why it matters
Raw hydrogen from an electrolyser is wet and carries some oxygen. Oxygen crossover is a safety issue, so the oxygen reading controls the deoxo reactor and detects faults in the stack. Downstream, water and oxygen are poisons for the ammonia synthesis catalyst, and fuel-cell vehicles need hydrogen to ISO 14687 Grade D.
In Saudi Arabia, NEOM Green Hydrogen Company's plant finished construction in August 2026 and is being commissioned, with commercial operation targeted for 2027. Projects of this kind need analyzers at every purification step and at the ammonia loop. Read the Saudi green hydrogen and ammonia guide.
Key limits
| Point | Typical value or limit |
|---|---|
| Raw hydrogen from an alkaline electrolyser | About 0.15 to 0.3 mol % O₂ (design study) |
| ISO 14687 Grade D: water | ≤ 5 µmol/mol, about −65 °C frost point at 1 atm |
| ISO 14687 Grade D: oxygen | ≤ 5 µmol/mol |
| ISO 14687 Grade D: other | CO₂ ≤ 2, CO ≤ 0.2, total sulfur ≤ 0.004 µmol/mol; hydrogen index ≥ 99.97 % |
| Ammonia synthesis make-up gas | Set by the catalyst licensor |
ISO 14687 limits are in µmol/mol, which equals ppm by volume. Check a set of results with the ISO 14687 hydrogen purity checker, and see the ISO 14687 hydrogen quality guide for the full list.
Where to measure
| # | Point | What is measured | YORK instrument |
|---|---|---|---|
| 1 | Raw hydrogen before the deoxo reactor | Oxygen in hydrogen (%) | PGA500-EX |
| 2 | After the deoxo reactor and TSA dryer | Trace oxygen and trace water | PGA500-EX, DPT990-EX |
| 3 | Nitrogen from the air separation unit | Trace oxygen and water in nitrogen | PGA500, DPT830 |
| 4 | Ammonia synthesis make-up gas (H₂ + N₂) | Trace water and oxygen | DPT990-EX, PGA500-EX |
| 5 | Refuelling dispenser | Water and oxygen against ISO 14687 Grade D | DPT990-EX, PGA500-EX |
YORK instruments cover water and oxygen. They do not measure hydrogen purity itself. Hydrogen in oxygen on the oxygen side of the stack, and impurities such as CO, CO₂ and sulfur at µmol/mol level, are covered by an integrated system supplied through YORK's partner integration factory.
Taking reliable readings
- Hazardous areas. Electrolyser halls, hydrogen compression and the ammonia loop are classified areas: use explosion-proof analyzers or an analyzer shelter. Certificates are provided per model with quotation.
- Sensor life in hydrogen. Hydrogen shortens the life of electrochemical oxygen sensors to about 12 months, so plan regular sensor replacement.
- Pressure. Reduce the sample to near-atmospheric pressure; YORK dew point analyzers accept up to 4 barg and convert to line-pressure dew point. Dispenser hydrogen needs a dedicated sampling system.
- Online and laboratory. Water and oxygen can be monitored online; the full ISO 14687 list is normally analysed in a laboratory.
Questions we are asked
What are the ISO 14687 Grade D limits for water and oxygen?
Both are at most 5 µmol/mol. For water this is about −65 °C frost point at atmospheric pressure. CO₂ is limited to 2 µmol/mol, CO to 0.2 µmol/mol and total sulfur to 0.004 µmol/mol, with a hydrogen index of at least 99.97 %.
Why measure oxygen in electrolytic hydrogen?
Raw hydrogen from an alkaline electrolyser can carry about 0.15 to 0.3 mol % oxygen. The reading controls the deoxo reactor and detects crossover, which is a safety issue.
Why do water and oxygen matter for green ammonia?
Both poison the ammonia synthesis catalyst, so the nitrogen and the make-up gas are dried and deoxygenated. The catalyst licensor sets the exact limits.
Which YORK analyzers suit hydrogen areas?
The explosion-proof DPT990-EX for dew point, from −110 °C frost point, and the explosion-proof PGA500-EX for trace oxygen, from 0.01 ppm to 25 % O₂. Certificates are provided per model with quotation. YORK does not supply a hydrogen purity analyzer of its own; that is part of an integrated system.
Suggested YORK instruments
- Trace water after the dryer, ammonia make-up gasDPT990-EX →
- Oxygen in hydrogen, raw and purifiedPGA500-EX →
- Nitrogen from the air separation unitPGA500 →
- Analyzer shelters and sample systemsAnalyzer panels →
See every point on the animated green hydrogen and refuelling measurement-point map.
Working on the nitrogen side? See ammonia, urea and air separation measurement points.