Process guide · animated
Ammonia, urea and air separation: where to measure moisture, O₂ and CO₂
Follow air through the air separation unit, and natural gas through the ammonia line to the urea reactor. Each numbered point shows what is measured, the typical limit and the analyzer that fits.
Tap a numbered point. Swipe the diagram sideways on a phone.
⤓ Download the 2-page PDF application note
How the process works
Gulf fertilizer complexes make ammonia from natural gas and turn most of it into urea with the CO₂ removed from the synthesis gas. Air separation units on the same sites supply oxygen and nitrogen. In both, catalysts and cryogenic equipment fail if trace water, CO₂ or oxygen gets through, so each purification step has an analyzer behind it.
- Air compression and front-end purification. Molecular-sieve adsorbers remove water and CO₂ from compressed air before it enters the cold box.
- Cryogenic separation. The cold box distils air into oxygen, nitrogen and argon. The reboiler sump is watched for hydrocarbons that could concentrate in liquid oxygen.
- Reforming and shift. Natural gas and steam are reformed to hydrogen and carbon monoxide; the shift converter turns most CO into CO₂ and more hydrogen.
- CO₂ removal and methanation. A solvent removes CO₂, which goes on to urea. The methanator converts the remaining CO and CO₂ to methane, because both poison the ammonia catalyst.
- Compression and ammonia synthesis. The hydrogen–nitrogen make-up gas is compressed into the synthesis loop, where it must be free of water and oxygen.
- Urea synthesis. CO₂, dosed with a little passivation air, is compressed and reacted with ammonia in the urea reactor.
Measurement points at a glance
| # | Point | Measures | Typical value / limit | Recommended solution |
|---|---|---|---|---|
| 1 | Front-end purification outlet | Trace CO₂ (ppmv) | Alarm typically at 1 ppmv; high-high ≤ 5 ppmv; shutdown at 10 ppmv | Integrated system: trace CO₂ NDIR analyzer |
| 2 | Front-end purification outlet | Water dew point | Detects adsorber breakthrough; −40 to −60 °C confirms the cold box is dry after deriming | DPT830, DPT810, DM9800, DM9900 |
| 3 | Main condenser / reboiler sump | Acetylene, N₂O and total hydrocarbons | Sump limits: C₂H₂ 0.5 ppmv; total hydrocarbons 500 ppmv as CH₄ | Integrated system: process gas chromatograph with flame-ionisation detector |
| 4 | Oxygen product | O₂ purity (%) | About 95 % for gasification; higher purity for chemical and merchant grades | POA3500, POA300, PGA530 |
| 5 | Nitrogen product | Trace O₂ (ppm) | Pure / high-purity / ultra-pure N₂: < 50 / < 3 / < 0.1 ppm (GB/T 8979-2025) | POA200, PGA500 |
| 6 | Nitrogen product | Trace water | < 15 / < 3 / < 0.5 ppm by grade (≈ −57 / −69 / −80 °C frost point) | DPT830, DM9900 |
| 7 | Methanator inlet and outlet | CO and CO₂ | Inlet typically about 0.2 % CO and 0.1 % CO₂; outlet < 5 ppmv each by design | Integrated system: NDIR CO/CO₂ analyzer at the inlet; gas chromatograph with methanizer-FID for ppm CO + CO₂ at the outlet |
| 8 | Synthesis make-up gas | Trace water | Water is a synthesis-catalyst poison; take the limit from the catalyst specification | DPT830, DPT990-EX |
| 9 | Synthesis make-up gas | Trace oxygen | Normally near zero; any oxygen forms water over the catalyst | PGA500-EX |
| 10 | Urea CO₂ feed | CO₂ purity (%) | > 98 % | Integrated system: CO₂ purity analyzer with a narrow high-percent range |
| 11 | Urea CO₂ feed after passivation air | Oxygen (vol %) | ≈ 0.6 vol % with conventional stainless steels; can be cut to ≈ 0.04 vol % with a special duplex stainless steel | POA3500; Integrated system: laser O₂ analyzer qualified for CO₂ service |
| 12 | Urea CO₂ feed | Hydrogen in CO₂ | Historically close to 1 %; many plants now remove it | Integrated system: thermal-conductivity H₂ analyzer |
Values are typical figures from published specifications and technical literature. Your contract, licensor or plant specification always governs.
Getting reliable readings
- Trace levels need tight sample systems. At ppm and sub-ppm levels, a small air leak or a wet fitting dominates the reading. Use stainless tubing, minimal dead volume and a steady purge.
- Oxygen service. Sample lines and wetted parts on oxygen product must be cleaned for oxygen service.
- Hydrogen-rich, high-pressure gas. Synthesis and make-up gas points are in hazardous areas. Reduce pressure in a certified sample system and use explosion-proof analyzers.
- CO₂ as the matrix. Electrochemical oxygen cells are not suited to a CO₂ background: for oxygen in the urea CO₂ feed use a paramagnetic or laser analyzer qualified for CO₂ service.
- Licensor limits govern. Catalyst and process licensors set the exact limits for each plant; the figures here are typical published values.
References
- AIGA 035/20 Safe Operation of Reboilers/Condensers in Air Separation UnitsAsia Industrial Gases Association
- Gasifipedia: Commercial OxygenUS DOE NETL
- GB/T 8979-2025 Pure nitrogen, high-purity nitrogen and ultra-pure nitrogen (table preview)chinesestandard.net
- Can passivation air in urea plants be minimized? IFA Technical Conference 2002International Fertilizer Association
- CN102659103A Preparation method of CO₂ feed gas for ureaGoogle Patents
- Gasifipedia: Water Gas ShiftUS DOE NETL
Related
- Air separation units: measurement points (animated)
- Saudi green hydrogen and ammonia: gas analysis guide
- All process maps by industry
- Industrial gas purity
- How to choose an oxygen analyzer
- Natural gas processing: measurement points (animated)
- CO₂ capture and utilisation: measurement points (animated)
- Oxygen analyzers
- Sampling and analyzer systems
Frequently asked questions
Why is CO₂ measured after the ASU front-end purification?
CO₂ that slips through the molecular-sieve adsorbers freezes in the cold box and can let hydrocarbons build up in liquid oxygen. Typical practice is an alarm at 1 ppmv and shutdown at 10 ppmv.
What oxygen level is used to passivate urea plants?
About 0.6 vol % oxygen in the CO₂ feed with conventional stainless steels. Plants built with a special duplex stainless steel can run at about 0.04 vol %.
Which points in an ammonia plant need a hydrogen-rated analyzer?
Synthesis make-up gas and other hydrogen-rich streams are hazardous areas. Moisture and oxygen there are measured with explosion-proof analyzers through a certified sample system.
Does YORK supply the GC and NDIR CO analyzers too?
Yes. Points such as methanator CO + CO₂ and reboiler hydrocarbons are supplied as an integrated system through YORK's partner integration factory, together with the YORK instruments.
Planning analyzers for an ammonia, urea or ASU plant?
Send us the process points, pressures and area classification, and we will propose the instruments and sampling systems.