Ex · Hazardous area
YORK PGA500-EX explosion-proof trace oxygen analyzer
The PGA500-EX is an explosion-proof, in-line trace oxygen analyzer, 0.01 ppm to 25% O₂, for installation in hazardous areas. It is used to monitor oxygen in nitrogen blanketing of hydrocarbon tanks and reactors, in inert gas supplies in process units, and in hydrogen, with outputs to the control system and alarm relays.

| Model | PGA500-EX |
|---|---|
| Type | Explosion-proof trace oxygen analyzer |
| Range | 0.01 ppm to 25% O₂ |
| Principle | Electrochemical |
| Explosion protection | Ex db IIC T6 Gb / Ex tb IIIC T80 °C Db; certificate details are sent with quotations |
| Housing | Explosion-proof aluminium alloy |
| Outputs | Analog output, digital output, alarm relay |
| Power | AC or DC supply |
| Not suitable for | CO₂ background and acid gases (H₂S, HCl, SO₂) |
| Hydrogen and hydrocarbons | Usable; sensor life is shorter, about 12 months |
| Lead time | 6 to 8 weeks |
| Warranty | 12 months from shipment |
| Regional contact | Dubai, UAE — covers the GCC |
Product catalog (PDF) · certificate copies are sent with quotations.
Where it is installed
| Location | Why oxygen is monitored |
|---|---|
| Nitrogen blanketing of hydrocarbon tanks and reactors | Keep oxygen below the safe limit; see process oxygen and inerting |
| Inert gas supply in process units | Oxygen in nitrogen used for purging and seals |
| Hydrogen production and refuelling | Oxygen in hydrogen; see green hydrogen measurement points |
| Process gases in refineries and petrochemical plants | Trace oxygen in nitrogen purge streams; see refinery measurement points |
Installation notes
- Use stainless steel sample lines and leak-tight fittings; at ppm levels any air leak shows in the reading.
- Keep a continuous sample flow.
- Do not use in CO₂ or acid gas service.
- In hydrogen and hydrocarbons, plan for cell replacement about every 12 months.
- YORK can build the sample system, panel or analyzer shelter around the analyzer.
PGA500-EX or another YORK analyzer?
| If you need… | Choose |
|---|---|
| Continuous trace O₂ in a hazardous area | PGA500-EX |
| Continuous trace O₂ in a safe area | PGA500 |
| Dew point in a hazardous area | DPT990-EX |
Ordering in the UAE and GCC
Send your enquiry to our Dubai team with:
- the gas and where it is measured;
- the expected oxygen range;
- whether the gas contains CO₂, acid gases (H₂S, HCl, SO₂), hydrogen or hydrocarbons;
- whether the measurement point is in a hazardous area;
- quantity.
We reply with a quotation for the PGA500-EX. Instruments are shipped from the factory with export documents, 6 to 8 weeks from order confirmation. Certificate details are sent with quotations. Where a project specifies IECEx or ATEX, ask us first; the analyzer can also be installed in a safe area with the sample line brought to it.
Questions about the PGA500-EX
What is the Ex marking of the PGA500-EX?
Ex db IIC T6 Gb / Ex tb IIIC T80 °C Db. Certificate details are sent with quotations.
What is the range of the PGA500-EX?
0.01 ppm to 25% O₂, electrochemical.
Can the PGA500-EX measure oxygen in hydrogen?
Yes; the electrochemical cell has a shorter life in hydrogen, about 12 months.
Can it measure oxygen in sour gas or CO₂?
No. CO₂ and acid gases shorten the cell life and affect the reading.
How do I get a quotation in the UAE?
Email dubai@yorkinstrument.com or WhatsApp +971 55 137 4866 with the gas, oxygen range, background components, area classification and quantity.