Refining and petrochemicals: where to measure moisture, O₂ and CO₂
Twelve points across hydrogen, reforming, isomerization, olefins, polymers, ethylene oxide and utilities. Each shows the typical limit, why it matters and the analyzer that fits.
Tap a numbered point. Swipe the diagram sideways on a phone.
How the process works
A refinery or petrochemical complex is a set of units, not one line. The map groups the units where gas analysis protects a catalyst, a product specification or plant safety.
- Hydrogen and hydrotreating. A PSA unit supplies make-up hydrogen. Trace CO and CO₂ in it inhibit hydrotreating catalysts.
- Catalytic reforming. The recycle hydrogen loop needs a tight water–chloride balance. Its HCl content rules out conventional metal-oxide moisture sensors.
- Isomerization. Feed driers protect a catalyst that water destroys permanently, so the drier outlet is alarmed.
- Steam cracking and olefins. Cracked gas is dried before the cold section. Polymer-grade ethylene and propylene are checked for water, oxygen, CO, CO₂ and acetylene before the PE/PP reactors.
- Ethylene oxide. The reactor cycle gas runs close to its flammable limit, so oxygen is part of the safety system and CO₂ is controlled for selectivity.
- Utilities and safety. Nitrogen blanketing keeps tanks below the limiting oxygen concentration, catalyst regeneration burns coke under controlled oxygen, and instrument air stays dry.
Measurement points at a glance
| # | Point | Measures | Typical value / limit | Recommended solution |
|---|---|---|---|---|
| 1 | Make-up hydrogen from the PSA unit | CO + CO₂ (ppm) | Typically < 10 ppmv; 10–50 ppmv slip is common; catalyst limits range from < 20 to 100 ppm | Integrated system: process GC with methanizer-FID |
| 2 | Catalytic reformer recycle hydrogen | Water and HCl | Water: semi-regen target 20 ± 5 ppmv; CCR 15–20 ppmv, max 30 ppmv. HCl: < 3 vol ppm | Integrated system: laser (TDLAS) moisture and HCl analyzer |
| 3 | Isomerization feed drier outlet | Trace water | Sub-ppm water: chlorided-alumina catalyst is irreversibly deactivated by water | DPT830, DPT990-EX |
| 4 | Cracked-gas molecular-sieve drier outlet | Trace water | Below 1 ppmv | Integrated system: laser (TDLAS) moisture analyzer |
| 5 | Polymer-grade ethylene / propylene | Water | Ethylene ≤ 7.8 ppm mol (5 ppmw), or ≤ 10 ppm v/v in another specification; propylene ≤ 5 ppm | DPT830, DPT990-EX |
| 6 | Polymer-grade ethylene / propylene | Oxygen | Ethylene ≤ 0.5 ppm mol; propylene ≤ 2–5 ppm | PGA500, PGA500-EX, POA200 |
| 7 | PE / PP reactor feed | CO, CO₂, acetylene | Ethylene: CO ≤ 1, CO₂ ≤ 3, C₂H₂ ≤ 3 ppm mol; propylene: CO ≤ 0.07–0.3 ppm | Integrated system: process GC with methanizer-FID |
| 8 | Ethylene oxide reactor cycle gas | Oxygen (%) | 5–10 mol % (8 % in a patent example); the shutdown trip is set from the flammability limit | POA3500; Integrated system: redundant paramagnetic oxygen analyzers (SIL-rated) |
| 9 | Ethylene oxide reactor cycle gas | CO₂ (%) | 1–15 mol % (another design 0.5–10 %) | Integrated system: NDIR CO₂ analyzer, explosion-proof or in a purged shelter |
| 10 | Tank and reactor nitrogen blanketing | Oxygen (%) | NFPA 69 margin, for example ≤ about 10 % O₂ for methane service and ≤ about 3 % for hydrogen | PGA500-EX |
| 11 | Catalyst regeneration gas | Oxygen (%) | 0.1–2.0 vol %; CCR burn zone usually 0.5–1.5 vol % | POA3500; Integrated system: paramagnetic or zirconia oxygen analyzer for acid-gas service |
| 12 | Instrument air header | Pressure dew point | At least 10 °C below the lowest ambient temperature, and never above 4 °C at line pressure (ISA-7.0.01) | DPT810, DM9200, DPT-600 Plus-EX |
Values are typical figures from published specifications and technical literature. Your contract, licensor or plant specification always governs.
Getting reliable readings
- Check the matrix first. HCl, heavy hydrocarbons and acid gases decide which sensor survives. Where a stream is chloride- or acid-bearing, YORK supplies an integrated analyzer suited to it.
- Vaporise liquids completely. Propylene and other liquefied streams must be vaporised without fractionation before a gas-phase measurement.
- Trace levels need tight sample systems. Sub-ppm water and oxygen need leak-tight, electropolished stainless tubing and steady flow.
- Hazardous area. Most process points are in Zone 1 or Zone 2: use explosion-proof analyzers or a certified analyzer shelter.
References
- Q&A 17: water content of reformer feed and recycle gasAFPM
- Q&A 44: CCR operating parameters including waterAFPM
- Q&A 53: typical CO/CO₂ in hydrotreating make-up hydrogenAFPM
- US 5053371: catalyst regeneration with three-zone combustion gas additionUSPTO
- US 8314281: light paraffin isomerization with improved feed purificationUSPTO
- Ethylene sales specificationChevron Phillips Chemical
- Polymerizable ethylene specificationORLEN
- Propylene specificationsIndian Oil
- SS-28 Propylene, polymer gradeRompetrol Rafinare
- Study of CO, CO₂ and O₂ traces as inhibitors of polypropylene synthesis, Polymers 2024MDPI
- US 6040467: high-purity oxygen for ethylene oxide productionGoogle Patents
- US 7851760: control and optimization of the ethylene oxide processUSPTO
- Inerting, purging and containment (NFPA 69 summary)Open Exam Prep
- US 4202865: on-line regeneration of hydrodesulfurization catalystUSPTO
- Understanding the ISA-7.0.01-1996 instrument air standardCompressed Air Best Practices
Frequently asked questions
Which moisture analyzer suits reformer recycle hydrogen?
The recycle gas carries HCl, which is corrosive to conventional metal-oxide moisture sensors. A laser (TDLAS) moisture analyzer is used, which YORK supplies as an integrated system.
What are typical limits for polymer-grade ethylene?
Published specifications set water at about 7.8 ppm mol, oxygen at 0.5 ppm mol, CO at 1 ppm mol, CO₂ at 3 ppm mol and acetylene at 3 ppm mol.
How much oxygen is allowed in a nitrogen-blanketed tank?
NFPA 69 keeps oxygen a margin below the limiting oxygen concentration, for example about 10 % for methane service and about 3 % for hydrogen.
What instrument air dew point is required?
ISA-7.0.01 asks for a pressure dew point at least 10 °C below the lowest ambient temperature, and never above 4 °C at line pressure.