Process guide · animated

Medical gases: where to measure moisture, O₂ and CO₂

Follow medical air from the compressors to the wards, and medical oxygen and CO₂ from storage to the pipeline and cylinders. Each point shows the pharmacopoeia limit and the analyzer that fits.

MEDICAL AIR TO THE WARDSMEDICAL O₂ PIPELINEMEDICAL AIR PLANTMEDICAL OXYGEN AND CO₂COMPRESSORSAFTERCOOLERRECEIVERDRYERSFILTERSLIQUID O₂VAPORISERPSA O₂O₂ MANIFOLDLIQUID CO₂CYLINDERFILLING1H₂O2O₂3CO₂4CO·SO₂·NOx5O₂ %6H₂O·CO₂7CO₂·H₂O
YORK instrumentIntegrated system (supplied through YORK)

Tap a numbered point. Swipe the diagram sideways on a phone.

Request a quote for this point →
Auto tour
Project list 0 →

⤓ Download the 2-page PDF application note

How the process works

Hospitals make medical air on site and receive oxygen and CO₂ as liquid or produce oxygen with PSA units. Pharmacopoeia monographs set the limits that each gas must meet before it reaches patients.

  1. Compression. Duty and standby compressors draw in air from a clean intake.
  2. Drying and filtration. Aftercoolers, desiccant dryers and filters remove water, oil and particles.
  3. Oxygen supply. Liquid oxygen is vaporised, or PSA units concentrate oxygen from air, and a manifold feeds the pipeline.
  4. Cylinders. Medical oxygen and CO₂ are filled into cylinders and released after testing.

Measurement points at a glance

#PointMeasuresTypical value / limitRecommended solution
1Medical air plant outletWater≤ 67 ppm (≈ −46 °C dew point at atmospheric pressure) in pharmacopoeia monographsDPT810, DM9600
2Medical air plant outletOxygen19.5–23.5 %POA3500, POA300
3Medical air plant outletCarbon dioxide≤ 500 ppmPGA520
4Medical air plant outletCO, SO₂, NO + NO₂, oilCO ≤ 5 ppm; SO₂ ≤ 1 ppm; NO + NO₂ ≤ 2 ppm; oil ≤ 0.1 ppmIntegrated system: electrochemical or NDIR CO, chemiluminescence NOx and oil-vapour monitoring
5Medical oxygen manifoldOxygen assay≥ 99.0 % for medical oxygenPOA3500
6Medical oxygen manifold / cylinder fillingWater and CO₂H₂O ≤ 67 ppm; CO₂ ≤ 300 ppmDPT830, DPT-600 Plus, PGA520
7Medical CO₂ cylinder fillingCO₂ purity and water≥ 99.0 %; H₂O ≤ 67 ppmDPT830; Integrated system: CO₂ assay analyzer (NDIR or GC) for the ≥ 99.0 % purity test

Values are typical figures from published specifications and technical literature. Your contract, licensor or plant specification always governs.

Practical notes

  • Pressure basis. The 67 ppm water limit corresponds to about −46 °C dew point at atmospheric pressure. At line pressure the dew point is higher, so convert before comparing.
  • Check the intake. CO₂ and CO at the plant outlet usually point to a poorly placed air intake.
  • Standards. Limits follow the pharmacopoeia monographs for medical gases; national hospital standards may add requirements.

Frequently asked questions

What is the water limit for medical air?

Pharmacopoeia monographs set water at 67 ppm or less, about −46 °C dew point at atmospheric pressure.

What oxygen content must medical air have?

Between 19.5 % and 23.5 %.

Which analyzers does a medical air plant need?

A dew point analyzer, an oxygen analyzer and a CO₂ analyzer, which YORK supplies, plus CO, SO₂, NOx and oil monitoring as an integrated system.

Equipping a medical gas plant?