Process guide · animated
CO₂ capture: where to measure O₂, moisture and CO₂ purity
Follow CO₂ from the flue gas through the amine unit, compression and drying to a pipeline or to beverage-grade storage. Each numbered point shows what is measured, the typical limit and the analyzer that fits.
Tap a numbered point. Swipe the diagram sideways on a phone.
⤓ Download the 2-page PDF application note
How the process works
Capture plants take CO₂ from flue gas or process off-gas, then compress and dry it for a pipeline or purify it for sale. The specification at the end decides which impurities need measuring along the way.
- Absorption. An amine solvent absorbs CO₂ from the flue gas in the absorber; the cleaned gas goes to the stack.
- Stripping. The rich amine is heated in the stripper, which releases nearly pure, water-saturated CO₂. The lean amine returns to the absorber.
- Compression. CO₂ is compressed in stages; water condenses out between the stages.
- Dehydration. A glycol contactor or molecular sieve dries the CO₂ so it cannot form carbonic acid or hydrates.
- Pipeline. Dense-phase CO₂ goes by pipeline to enhanced oil recovery (EOR) or storage, against limits on water, oxygen and other impurities.
- Beverage grade. Part of the CO₂ can be purified and liquefied to meet the ISBT beverage-grade guidelines.
Measurement points at a glance
| # | Point | Measures | Typical value / limit | Recommended solution |
|---|---|---|---|---|
| 1 | Absorber inlet (flue gas) | Oxygen | Percent level in combustion flue gas | POA3500; Integrated system: in-situ zirconia O₂ probe for the hot, wet gas |
| 2 | Stripper overhead | CO₂ purity (%) | High-purity, water-saturated CO₂ on its way to the ≥ 95 % product specification | PGA520, PGA620 |
| 3 | Dehydration outlet | Water content | 500 ppmv design limit for EOR pipelines (≈ −27 °C frost point at 1 atm) | DPT830, DPT810, DM9700 |
| 4 | CO₂ pipeline to EOR / storage | Trace oxygen | 0.001 vol % (10 ppmv) design value | Integrated system: trace O₂ analyzer qualified for CO₂ service |
| 5 | CO₂ pipeline to EOR / storage | CO₂ purity | ≥ 95 % CO₂ (design value) | PGA520 |
| 6 | CO₂ pipeline to EOR / storage | CO, H₂S, N₂ and other impurities | Design values: CO 35 ppmv, H₂S 0.01 vol %, N₂ 1 % | Integrated system: process gas chromatograph with sulfur-selective detection |
| 7 | Beverage-grade CO₂ storage | Moisture | ≤ 20 ppm v/v under the ISBT guidelines (≈ −55 °C frost point at 1 atm) | DPT810, DM9800 |
| 8 | Beverage-grade CO₂ storage | Trace oxygen | ≤ 30 ppm v/v under the ISBT guidelines | Integrated system: trace O₂ analyzer qualified for CO₂ service |
| 9 | Beverage-grade CO₂ product | CO₂ purity | ≥ 99.9 % under the ISBT guidelines | Integrated system: high-resolution CO₂ purity analysis (GC or high-precision NDIR) |
| 10 | Beverage-grade CO₂ product | CO, total sulfur, hydrocarbons, benzene | ISBT limits include CO ≤ 10 ppm v/v and total sulfur ≤ 0.1 ppm v/v | Integrated system: gas chromatograph with sulfur-selective detection, or FTIR |
Values are typical figures from published specifications and technical literature. Your contract, licensor or plant specification always governs.
Getting reliable readings
- Pressure basis. Dense-phase CO₂ is measured after pressure let-down; always record at what pressure a dew point applies.
- CO₂ matrix. Electrochemical oxygen cells are not suited to nearly pure CO₂: use an oxygen analyzer qualified for CO₂ service.
- Vapour phase. For liquid CO₂, take the sample from the vapour phase after a vaporizer, or follow the method of your quality specification.
- Specifications. EOR and storage limits are set per project; the NETL values used here are design parameters, not a universal standard.
References
- QGESS: CO₂ Impurity Design Parameters (NETL-PUB-22529)US DOE National Energy Technology Laboratory
- Gasifipedia: Acid Gas RemovalUS DOE National Energy Technology Laboratory
Related
- Dates: storage and packing measurement points (animated)
- Gulf CCS: CO₂ dehydration and moisture measurement
- All process maps by industry
- CO₂ in food, beverage and MAP
- How an NDIR CO₂ analyzer works (interactive)
- Natural gas processing: measurement points (animated)
- Ammonia, urea and air separation: measurement points (animated)
- CO₂ analyzers
Frequently asked questions
What water content is allowed in a CO₂ pipeline?
NETL design parameters for EOR pipelines use 500 ppmv water. The project or pipeline operator sets the actual limit.
What are the ISBT limits for beverage-grade CO₂?
The ISBT guidelines include purity of at least 99.9 %, moisture of no more than 20 ppm v/v and oxygen of no more than 30 ppm v/v, plus limits on CO, sulfur and hydrocarbons.
Why measure oxygen at the absorber inlet?
Oxygen in the flue gas oxidizes and degrades the amine solvent, so it affects solvent consumption.
Can one supplier cover all the analyzers?
YORK supplies moisture, oxygen and CO₂ instruments and, through its partner integration factory, integrated systems for GC and sulfur analysis.
Planning analyzers for a CO₂ project?
Send us the process points, pressures and product specification, and we will propose the instruments and sampling systems.