CO₂ capture: where to measure O₂, moisture and CO₂ purity

Follow CO₂ from the flue gas through the amine unit, compression and drying to a pipeline or to beverage-grade storage. Each numbered point shows what is measured, the typical limit and the analyzer that fits.

FLUE GASTO STACKTO EOR / STORAGELEAN AMINERICH AMINEABSORBERSTRIPPERCO₂ COMPRESSORSTAGE 2DEHYDRATIONDENSE-PHASE PUMPPURIFICATION +LIQUEFACTIONBEVERAGE CO₂1O₂2CO₂3H₂O4O₂5CO₂6GC7H₂O8O₂9CO₂10GC
YORK instrumentIntegrated system (supplied through YORK)

Tap a numbered point. Swipe the diagram sideways on a phone.

Request a quote for this point →
Auto tour
Project list 0 →

⤓ Download the 2-page PDF application note

How the process works

Capture plants take CO₂ from flue gas or process off-gas, then compress and dry it for a pipeline or purify it for sale. The specification at the end decides which impurities need measuring along the way.

  1. Absorption. An amine solvent absorbs CO₂ from the flue gas in the absorber; the cleaned gas goes to the stack.
  2. Stripping. The rich amine is heated in the stripper, which releases nearly pure, water-saturated CO₂. The lean amine returns to the absorber.
  3. Compression. CO₂ is compressed in stages; water condenses out between the stages.
  4. Dehydration. A glycol contactor or molecular sieve dries the CO₂ so it cannot form carbonic acid or hydrates.
  5. Pipeline. Dense-phase CO₂ goes by pipeline to enhanced oil recovery (EOR) or storage, against limits on water, oxygen and other impurities.
  6. Beverage grade. Part of the CO₂ can be purified and liquefied to meet the ISBT beverage-grade guidelines.

Measurement points at a glance

#PointMeasuresTypical value / limitRecommended solution
1Absorber inlet (flue gas)OxygenPercent level in combustion flue gasPOA3500; Integrated system: in-situ zirconia O₂ probe for the hot, wet gas
2Stripper overheadCO₂ purity (%)High-purity, water-saturated CO₂ on its way to the ≥ 95 % product specificationPGA520, PGA620
3Dehydration outletWater content500 ppmv design limit for EOR pipelines (≈ −27 °C frost point at 1 atm)DPT830, DPT810, DM9700
4CO₂ pipeline to EOR / storageTrace oxygen0.001 vol % (10 ppmv) design valueIntegrated system: trace O₂ analyzer qualified for CO₂ service
5CO₂ pipeline to EOR / storageCO₂ purity≥ 95 % CO₂ (design value)PGA520
6CO₂ pipeline to EOR / storageCO, H₂S, N₂ and other impuritiesDesign values: CO 35 ppmv, H₂S 0.01 vol %, N₂ 1 %Integrated system: process gas chromatograph with sulfur-selective detection
7Beverage-grade CO₂ storageMoisture≤ 20 ppm v/v under the ISBT guidelines (≈ −55 °C frost point at 1 atm)DPT810, DM9800
8Beverage-grade CO₂ storageTrace oxygen≤ 30 ppm v/v under the ISBT guidelinesIntegrated system: trace O₂ analyzer qualified for CO₂ service
9Beverage-grade CO₂ productCO₂ purity≥ 99.9 % under the ISBT guidelinesIntegrated system: high-resolution CO₂ purity analysis (GC or high-precision NDIR)
10Beverage-grade CO₂ productCO, total sulfur, hydrocarbons, benzeneISBT limits include CO ≤ 10 ppm v/v and total sulfur ≤ 0.1 ppm v/vIntegrated system: gas chromatograph with sulfur-selective detection, or FTIR

Values are typical figures from published specifications and technical literature. Your contract, licensor or plant specification always governs.

Getting reliable readings

  • Pressure basis. Dense-phase CO₂ is measured after pressure let-down; always record at what pressure a dew point applies.
  • CO₂ matrix. Electrochemical oxygen cells are not suited to nearly pure CO₂: use an oxygen analyzer qualified for CO₂ service.
  • Vapour phase. For liquid CO₂, take the sample from the vapour phase after a vaporizer, or follow the method of your quality specification.
  • Specifications. EOR and storage limits are set per project; the NETL values used here are design parameters, not a universal standard.

References

  1. QGESS: CO₂ Impurity Design Parameters (NETL-PUB-22529)US DOE National Energy Technology Laboratory
  2. Gasifipedia: Acid Gas RemovalUS DOE National Energy Technology Laboratory

Frequently asked questions

What water content is allowed in a CO₂ pipeline?

NETL design parameters for EOR pipelines use 500 ppmv water. The project or pipeline operator sets the actual limit.

What are the ISBT limits for beverage-grade CO₂?

The ISBT guidelines include purity of at least 99.9 %, moisture of no more than 20 ppm v/v and oxygen of no more than 30 ppm v/v, plus limits on CO, sulfur and hydrocarbons.

Why measure oxygen at the absorber inlet?

Oxygen in the flue gas oxidizes and degrades the amine solvent, so it affects solvent consumption.

Can one supplier cover all the analyzers?

YORK supplies moisture, oxygen and CO₂ instruments and, through its partner integration factory, integrated systems for GC and sulfur analysis.