Guide · calibration
How to calibrate a dew point sensor
Dew point sensors drift. Metal-oxide and polymer sensors slowly change as they age, as contaminants build up on the sensor surface, and after exposure to liquid water or aggressive gases. Calibration checks the sensor against a reference of known accuracy at one or more dew points and, where needed, adjusts it. This guide explains the two ways it is done, how often, and what equipment a laboratory or a site team needs.
At a glance
| Typical interval | Every 12 months for most process transmitters; shorter in dirty or wet gas, or where the reading is used for custody transfer or product release |
|---|---|
| Field check | Compare against a portable reference at the process dew point; quick, one point |
| Laboratory calibration | Several dew points across the working range, using a humidity generator and a chilled mirror reference |
| Reference instrument | A chilled mirror hygrometer, because it measures dew point directly |
| Traceability | Reference calibrated by an ISO/IEC 17025 laboratory traceable to a national metrology institute |
Why dew point sensors drift
- Metal-oxide sensors respond to water vapour adsorbed in the pores of the oxide film. Over months, the pore structure changes slightly and the calibration shifts, typically by a few degrees at low dew points.
- Oil mist, glycol, methanol and particles coat the sensor and slow its response.
- Liquid water, or a large step from a very dry to a wet gas, can shift the calibration until the sensor has been dried out and checked again.
- Acid gases such as HCl and H₂S can damage the sensor permanently.
A sensor that reads correctly at −20 °C can still be wrong at −60 °C, which is why low dew point applications need checks near the working point.
Method 1: field check against a portable reference
- Connect a portable reference, such as a chilled mirror hygrometer or a recently calibrated portable dew point meter, to the same sample point, ideally through a tee so that both see the same gas.
- Use stainless steel tubing and keep the line short. Plastic tubing lets moisture in and makes low readings slow and too wet.
- Wait for both readings to stabilise. At −60 °C this can take an hour or more; at −20 °C, a few minutes.
- Compare at the same pressure. If the transmitter reads at line pressure and the reference at atmospheric pressure, convert one reading first with the dew point calculator.
- Record both readings, the difference and the acceptance limit. If the difference is outside the limit, send the transmitter for laboratory calibration or replace the sensor.
A field check verifies one point. It does not replace a multi-point calibration, but it catches a failed sensor between calibrations.
Method 2: multi-point laboratory calibration
- Generate a stable dew point with a low humidity generator, starting from a dry gas source.
- Feed the same gas to the reference hygrometer and to the sensors under test.
- Step through several dew points that cover the working range, for example −60, −40, −20 and 0 °C for a compressed air transmitter, or down to −80 °C for a sensor used after molecular sieve dryers. Work from dry to wet, because drying a system takes much longer than wetting it.
- At each point, wait until the generator, the reference and the sensors are stable, then record the readings.
- Adjust the sensor if the instrument allows it, repeat the check, and issue a certificate that lists each point, the reference reading, the error and the uncertainty.
Equipment for a calibration bench
| Item | What it does | YORK option |
|---|---|---|
| Dry gas source | Supplies gas far drier than the lowest calibration point | DC1000: dry air below 1 ppmV (−75 °C dew point), up to 10 NL/min |
| Low humidity generator | Mixes dry and wet gas to stable set points | DC2000: stability better than ±0.2 °C frost point, three gas outlets |
| Reference hygrometer | Measures the generated dew point directly | DM9900: −90 °C frost point to +20 °C, better than ±0.2 °C near −90 °C |
| Complete system | Compressor, dry gas generator, humidity generator and reference in a 19-inch rack | YKDC220 |
| Field reference | Portable checks at site | DM9600P portable chilled mirror, or DPT-600 Plus |
Uncertainty and traceability
The result can be no better than the reference. A chilled mirror reference is itself calibrated by an ISO/IEC 17025 laboratory, which is traceable to a national metrology institute. In the UAE, laboratories are accredited by EIAC; in Saudi Arabia, by SAAC. When you compare offers for calibration, ask for the calibration points, the uncertainty at each point and the accreditation scope, not only the price.
A calibration certificate for the DM9900 is available from YORK (chargeable). For building a laboratory bench, see dew point calibration laboratories.
Common mistakes
- Calibrating only at one point near ambient for a sensor that works at −60 °C.
- Not waiting long enough at low dew points.
- Plastic tubing or leaking fittings between the generator and the sensors.
- Comparing a line-pressure reading with an atmospheric-pressure reading without converting.
- Calibrating a sensor that is still wet from a process upset.
Related
Frequently asked questions
How often should a dew point sensor be calibrated?
Most process dew point transmitters are calibrated every 12 months. Shorten the interval where the gas carries oil, glycol or methanol, after a liquid water upset, or where the reading is used for product release or custody transfer. A field check between calibrations catches a failed sensor early.
Can I calibrate a dew point transmitter on site?
You can verify it on site against a portable reference at the process dew point. A full calibration with several points and a certificate is done in a laboratory with a humidity generator and a chilled mirror reference.
What reference is used to calibrate dew point sensors?
A chilled mirror hygrometer, because it measures the dew point directly. The reference itself is calibrated by an ISO/IEC 17025 laboratory traceable to a national metrology institute.