YORK DM9600PL portable chilled mirror hygrometer

The DM9600PL is a battery-powered chilled mirror hygrometer for site work, from −70 °C frost point to +20 °C dew point. It shares the housing and features of the DM9600P but reaches 10 °C lower, so it is chosen for field checks of desiccant dryers, nitrogen and other dry, non-flammable gases below −60 °C.

YORK DM9600PL – Portable chilled mirror hygrometer
The DM9600PL shares its housing with the DM9600P; only the nameplate differs.
ModelDM9600PL
TypePortable chilled mirror hygrometer
Range−70 °C frost point to +20 °C dew point
PrincipleChilled mirror, air cooled
Built inPressure sensor, electronic flow meter
Data and powerData logging to SD card; rechargeable lithium battery
UseNon-flammable gases in safe areas; sample pressure up to 4 barg
Lead time6 to 8 weeks
Warranty12 months from shipment
Regional contactDubai, UAE — covers the GCC

Product catalog (PDF) · certificate copies are sent with quotations.

Where it is used

JobWhat you check
Desiccant dryer outletsDew point below −60 °C at the outlet; see compressed air dew point
Nitrogen and argon suppliesMoisture at the point of use; see industrial gas purity
SF₆ and SF₆-free switchgearMoisture in the insulating gas; see the SF₆ moisture calculator
Verifying fixed transmittersSide-by-side check of low dew point transmitters at site

Good practice on site

  • Reduce the sample to 4 barg or less and use short stainless steel lines.
  • At −60 °C and below, allow longer for the lines and mirror to stabilise.
  • Record only when the frost point reading is stable; use SD logging to keep the record.
  • YORK chilled mirror hygrometers are low-pressure instruments for non-flammable gases in safe areas; they are not explosion-proof and are not intended for natural gas, hydrogen or other flammable gases.

DM9600PL or another YORK chilled mirror?

If you need…Choose
High dew points up to +90 °C, climatic chambers, insertion probeDM9200
Benchtop instrument to −60 °C frost pointDM9600
Portable checks to −60 °C frost pointDM9600P
Rack-mount monitoring to −70 or −80 °C frost pointDM9700 or DM9800
Laboratory reference to −90 °C frost pointDM9900
Flammable gases or hazardous areasA metal-oxide instrument: DPT-600 Plus-EX or DPT990-EX

Ordering in the UAE and GCC

Send your enquiry to our Dubai team with:

  • the gas and where it is measured;
  • the expected dew point and pressure;
  • quantity.

We reply with a quotation for the DM9600PL. Instruments are shipped from the factory with export documents, 6 to 8 weeks from order confirmation. The DM9600PL is battery powered for site use in safe areas.

See all dew point meters and analyzers available in the UAE.

Questions about the DM9600PL

What is the range of the DM9600PL?

−70 °C frost point to +20 °C dew point.

What is the difference between the DM9600PL and the DM9600P?

They share the same housing and features; the DM9600PL reaches −70 °C frost point, 10 °C lower than the DM9600P.

Can the DM9600PL measure SF₆?

Yes. Convert the dew point to ppmv with the SF₆ moisture calculator and compare with the equipment limit.

Can it be used for natural gas or other flammable gases?

No. YORK chilled mirror hygrometers are low-pressure instruments (up to 4 barg) and are not explosion-proof; they are for non-flammable gases in safe areas. For natural gas, see the DPT990-EX.

How do I get a quotation in the UAE?

Email dubai@yorkinstrument.com or WhatsApp +971 55 137 4866 with the gas, dew point range and quantity.