Guide · technology comparison
Chilled mirror, laser (TDLAS) or metal-oxide: choosing a moisture analyzer
Three technologies cover most moisture measurements in gases. A chilled mirror hygrometer measures the dew point directly by cooling a mirror until water condenses on it. A metal-oxide sensor measures the water vapour adsorbed in a thin oxide film. A tunable diode laser absorption spectroscopy (TDLAS) analyzer measures how much laser light water molecules absorb at a specific wavelength. Each is the right choice somewhere and the wrong one elsewhere. This guide explains where.
At a glance
| Chilled mirror | Metal-oxide | Laser (TDLAS) | |
|---|---|---|---|
| What it measures | Dew or frost point, directly | Water vapour, via a sensor calibrated against a reference | Water concentration from light absorption |
| Typical role | Reference, calibration laboratories, critical checks | Process transmitters and portable meters; wide range at moderate cost | Natural gas with contaminants; fast response; no sensor in contact with the gas |
| Response at low dew points | Slow; the mirror must reach and hold the frost point | Fast to moderate; slower at trace levels | Fast |
| Drift | Very low; the measurement is fundamental | Drifts with time; needs regular calibration | Low; checked against a validation gas |
| Sensitive to | Contamination on the mirror; condensing hydrocarbons in natural gas | Glycol, methanol, oil, liquid water, acid gases | Changes in background gas composition, which the analyzer must be set up for |
| Cost | Moderate to high | Lowest | Highest |
| From YORK | DM series, made by YORK | DPT series, made by YORK | Supplied as an integrated system |
How each one works
Chilled mirror. A small mirror is cooled until a thin layer of dew or frost forms on it. An optical system keeps the layer at constant thickness, and the mirror temperature at that point is the dew or frost point. Because it measures the condensation temperature itself, it is used as a reference. See how a chilled mirror hygrometer works.
Metal-oxide. Water vapour diffuses into the pores of a thin aluminium oxide film between two electrodes and changes its capacitance. The sensor is calibrated against a reference at several dew points. It is small, robust and covers a wide range, down to −110 °C frost point in the DPT series.
Laser (TDLAS). A laser is tuned across a water absorption line, and the analyzer measures how much light is absorbed along a fixed path through the sample. The gas only touches the windows of the sample cell, not a sensor. The absorption depends on the background gas, so the analyzer is set up and validated for the gas composition it will see.
Where each one fits
| Application | Usual choice | Why |
|---|---|---|
| Calibration laboratory reference | Chilled mirror | Direct measurement of dew point, traceable calibration |
| Instrument air and compressed air | Metal-oxide | Clean gas, moderate cost, many points to measure |
| Industrial gas purity (nitrogen, argon, oxygen) | Metal-oxide, with chilled mirror for checks | Clean, dry gas; trace range needed |
| Natural gas water dew point, clean dry gas | Metal-oxide | Cost-effective with good sample conditioning |
| Natural gas with glycol or methanol carry-over, or with H₂S | TDLAS | No sensor in contact with the contaminants; fast response |
| Reformer recycle hydrogen with HCl | TDLAS | HCl is corrosive to metal-oxide sensors |
| Molecular sieve dryer outlet | Metal-oxide or TDLAS | Trace range and fast response; see molecular sieve breakthrough |
| SF₆ and SF₆-free switchgear | Chilled mirror or metal-oxide portable | Field checks against the equipment's moisture limit |
| Hydrocarbon dew point | None of these three | Needs a dedicated hydrocarbon dew point analyzer |
Natural gas: the special case for chilled mirrors
In natural gas, heavier hydrocarbons can condense on the mirror at a temperature close to, or above, the water dew point. A manual or basic automatic chilled mirror can then report the hydrocarbon dew point instead of the water dew point. This is why process water dew point measurement in natural gas usually uses metal-oxide or TDLAS analyzers, and chilled mirrors are used there mainly by experienced operators for checks. Standards such as ISO 6327 and ASTM D1142 describe the chilled mirror method for natural gas. YORK chilled mirror hygrometers are low-pressure instruments for laboratories and safe areas; they are not explosion-proof and are not intended for natural gas on site. See also natural gas water dew point.
How to choose
- Start with the gas. Clean, dry gas suits any of the three. Gas with glycol, methanol, amine, H₂S or HCl points to TDLAS.
- Then the job. A reference or calibration measurement points to a chilled mirror. Many process points with a limited budget point to metal-oxide.
- Then the range and response. Trace levels with fast changes, such as dryer breakthrough, need a fast analyzer and a short, clean sample line, whatever the technology.
- Then the site. In a hazardous area, use an explosion-proof analyzer or put the analyzer in an analyzer shelter. Certificate details are sent with quotations.
- Then the maintenance. Metal-oxide sensors need calibration, typically every 12 months; see how to calibrate a dew point sensor. Chilled mirrors need the mirror cleaned. TDLAS analyzers are checked against a validation gas.
YORK options
| Technology | YORK instruments |
|---|---|
| Chilled mirror | DM9900 laboratory reference (−90 °C frost point to +20 °C, better than ±0.2 °C near −90 °C), DM9700, DM9600P portable |
| Metal-oxide | DPT810, DPT830 in-line; DPT990-EX explosion-proof; DPT-600 Plus portable; down to −110 °C frost point |
| Laser (TDLAS) | Supplied as an integrated system, with the sample system, panel or analyzer shelter built by YORK in Chengdu |
Related
Frequently asked questions
Is TDLAS better than a chilled mirror?
Neither is better in general. TDLAS suits natural gas with contaminants and fast changes, because no sensor touches the gas. A chilled mirror measures the dew point directly and is the usual reference in calibration laboratories.
Why are chilled mirrors not common for process water dew point in natural gas?
Heavier hydrocarbons can condense on the mirror near or above the water dew point, so the reading can be the hydrocarbon dew point instead. Process measurements therefore usually use metal-oxide or TDLAS analyzers.
Do metal-oxide moisture sensors work in natural gas?
Yes, in clean, dry gas with good sample conditioning: a pressure reducer, filters and a short stainless steel line. Glycol, methanol, liquid water and acid gases shorten sensor life and cause drift; where carry-over is persistent, TDLAS is the alternative.
Does YORK supply TDLAS moisture analyzers?
Yes, as an integrated system. YORK builds the sample system, panel or analyzer shelter around the analyzer at its factory in Chengdu.