Air separation units: where to measure moisture and oxygen

Follow the air from the intake through the front-end adsorbers into the cold box, and out as oxygen, nitrogen and argon products. Each numbered point shows what is measured, the typical value, why it matters and the instrument that fits.

FRONT ENDON-SITE PSA NITROGENCOLD BOXPRODUCTS AND STORAGEINTAKECOMPRESSORCOOLERADSORBERSMAIN EXCHANGERLPHPCRUDE ArPURE ArLOXLINLARARGONNITROGENOXYGENCYLINDER FILLING1H₂O2CO₂3C₂H₂4O₂5O₂6O₂ %7O₂8H₂O9O₂
YORK instrumentIntegrated system (supplied through YORK)

Tap a numbered point. Swipe the diagram sideways on a phone.

Request a quote for this point →
Auto tour
Project list 0 →

How the process works

Air separation supplies oxygen, nitrogen and argon to refineries, steel and aluminium plants, gas-to-liquids, hospitals and electronics. Moisture and CO₂ protect the cold box; oxygen purity and trace oxygen define the products.

  1. Compression and cooling. The main air compressor delivers air to a direct contact cooler, which removes heat and some water.
  2. Front-end purification. Two adsorbers in turn remove water, CO₂ and most hydrocarbons; one is regenerated while the other is on line.
  3. Cold box. The main heat exchanger cools the air close to its liquefaction point, and the high- and low-pressure columns separate it into oxygen and nitrogen. The main condenser links the two columns.
  4. Argon. A crude argon column draws an argon-rich stream from the low-pressure column; a pure argon column or a deoxo unit removes the remaining oxygen and nitrogen.
  5. Products. Gaseous products go to the pipeline; liquid oxygen, nitrogen and argon go to storage tanks, tankers and cylinder filling.
  6. On-site generators. Smaller users often make nitrogen on site with PSA or membrane generators instead of buying liquid.

Measurement points at a glance

#PointMeasuresTypical value / limitRecommended solution
1Front-end adsorber outletWater dew pointDetects adsorber breakthrough; −40 to −60 °C confirms the cold box is dry after derimingDPT830, DPT810, DM9900
2Front-end adsorber outletTrace CO₂ (ppmv)Alarm typically at 1 ppmv; high-high ≤ 5 ppmv; shutdown at 10 ppmvTrace CO₂ NDIR analyzer
3Main condenser sumpAcetylene, N₂O and total hydrocarbonsSump limits: C₂H₂ 0.5 ppmv; total hydrocarbons 500 ppmv as CH₄Process gas chromatograph with flame-ionisation detector
4Crude argon columnOxygen in crude argonPercent-level oxygen from a conventional sidearm column (one design targets ≤ 0.5 mol %); ppm-level in plants with a fully cryogenic argon columnPGA530, PGA500
5Pure argon productTrace oxygen (and nitrogen)Argon 5.0 (99.999 %) lists about 2 ppm O₂, 5 ppm N₂ and 3 ppm H₂OPGA500, POA200
Integrated system: nitrogen-in-argon analyzer
6Oxygen product and LOXOxygen purity (%)Medicinal oxygen ≥ 99.5 % (Ph. Eur.); about 95 % for gasification; industrial grades set by contractPOA3500, PGA530, POA300
7Nitrogen product and LINTrace oxygen (ppm)Pure / high-purity / ultra-pure N₂: < 50 / < 3 / < 0.1 ppm (GB/T 8979-2025)PGA500, POA200
8Cylinder filling manifoldWater in O₂, N₂ and ArMedicinal oxygen ≤ 67 ppm H₂O; nitrogen < 15 / < 3 / < 0.5 ppm by grade; argon 5.0 about 3 ppmDPT-600 Plus, DPT830
9On-site PSA nitrogen generatorResidual oxygen in product nitrogenPSA units reach ppm-level oxygen; membrane units leave percent-level oxygen. The set point comes from the user's processPGA530, POA300, PGA500

Values are typical figures from published specifications and technical literature. Your contract, licensor or plant specification always governs.

Getting reliable readings

  • Protect the cold box first. Treat the adsorber outlet water and CO₂ analyzers as safety instruments, with alarms and regular checks.
  • Match the range to the stream. Crude argon and on-site nitrogen may need percent or ppm ranges depending on the design.
  • Oxygen service. Use oxygen-clean sample systems on oxygen-rich streams.
  • Check the filling line. Cylinder water problems often come from the manifold, not from the plant.

Frequently asked questions

Where is moisture measured in an air separation unit?

At the outlet of the front-end adsorbers, to catch breakthrough before water reaches the cold box, and in the products and at cylinder filling as part of the product specification.

How is oxygen purity measured in an ASU?

Percent oxygen in the oxygen product is measured with a paramagnetic or electrochemical analyzer; trace oxygen in nitrogen and argon with a ppm-range electrochemical analyzer.

Is CO₂ at the adsorber outlet a safety issue?

Yes. CO₂ that reaches the cold box can plug the reboiler and let hydrocarbons concentrate in liquid oxygen, so plants alarm at about 1 ppmv.

Planning or upgrading an ASU?

Send us the products, purities and sample points, and we will propose the analyzers and sample systems.