Air separation units: where to measure moisture and oxygen
Follow the air from the intake through the front-end adsorbers into the cold box, and out as oxygen, nitrogen and argon products. Each numbered point shows what is measured, the typical value, why it matters and the instrument that fits.
Tap a numbered point. Swipe the diagram sideways on a phone.
How the process works
Air separation supplies oxygen, nitrogen and argon to refineries, steel and aluminium plants, gas-to-liquids, hospitals and electronics. Moisture and CO₂ protect the cold box; oxygen purity and trace oxygen define the products.
- Compression and cooling. The main air compressor delivers air to a direct contact cooler, which removes heat and some water.
- Front-end purification. Two adsorbers in turn remove water, CO₂ and most hydrocarbons; one is regenerated while the other is on line.
- Cold box. The main heat exchanger cools the air close to its liquefaction point, and the high- and low-pressure columns separate it into oxygen and nitrogen. The main condenser links the two columns.
- Argon. A crude argon column draws an argon-rich stream from the low-pressure column; a pure argon column or a deoxo unit removes the remaining oxygen and nitrogen.
- Products. Gaseous products go to the pipeline; liquid oxygen, nitrogen and argon go to storage tanks, tankers and cylinder filling.
- On-site generators. Smaller users often make nitrogen on site with PSA or membrane generators instead of buying liquid.
Measurement points at a glance
| # | Point | Measures | Typical value / limit | Recommended solution |
|---|---|---|---|---|
| 1 | Front-end adsorber outlet | Water dew point | Detects adsorber breakthrough; −40 to −60 °C confirms the cold box is dry after deriming | DPT830, DPT810, DM9900 |
| 2 | Front-end adsorber outlet | Trace CO₂ (ppmv) | Alarm typically at 1 ppmv; high-high ≤ 5 ppmv; shutdown at 10 ppmv | Trace CO₂ NDIR analyzer |
| 3 | Main condenser sump | Acetylene, N₂O and total hydrocarbons | Sump limits: C₂H₂ 0.5 ppmv; total hydrocarbons 500 ppmv as CH₄ | Process gas chromatograph with flame-ionisation detector |
| 4 | Crude argon column | Oxygen in crude argon | Percent-level oxygen from a conventional sidearm column (one design targets ≤ 0.5 mol %); ppm-level in plants with a fully cryogenic argon column | PGA530, PGA500 |
| 5 | Pure argon product | Trace oxygen (and nitrogen) | Argon 5.0 (99.999 %) lists about 2 ppm O₂, 5 ppm N₂ and 3 ppm H₂O | PGA500, POA200 Integrated system: nitrogen-in-argon analyzer |
| 6 | Oxygen product and LOX | Oxygen purity (%) | Medicinal oxygen ≥ 99.5 % (Ph. Eur.); about 95 % for gasification; industrial grades set by contract | POA3500, PGA530, POA300 |
| 7 | Nitrogen product and LIN | Trace oxygen (ppm) | Pure / high-purity / ultra-pure N₂: < 50 / < 3 / < 0.1 ppm (GB/T 8979-2025) | PGA500, POA200 |
| 8 | Cylinder filling manifold | Water in O₂, N₂ and Ar | Medicinal oxygen ≤ 67 ppm H₂O; nitrogen < 15 / < 3 / < 0.5 ppm by grade; argon 5.0 about 3 ppm | DPT-600 Plus, DPT830 |
| 9 | On-site PSA nitrogen generator | Residual oxygen in product nitrogen | PSA units reach ppm-level oxygen; membrane units leave percent-level oxygen. The set point comes from the user's process | PGA530, POA300, PGA500 |
Values are typical figures from published specifications and technical literature. Your contract, licensor or plant specification always governs.
Getting reliable readings
- Protect the cold box first. Treat the adsorber outlet water and CO₂ analyzers as safety instruments, with alarms and regular checks.
- Match the range to the stream. Crude argon and on-site nitrogen may need percent or ppm ranges depending on the design.
- Oxygen service. Use oxygen-clean sample systems on oxygen-rich streams.
- Check the filling line. Cylinder water problems often come from the manifold, not from the plant.
References
- AIGA 035/20 Safe Operation of Reboilers/Condensers in Air Separation UnitsAsia Industrial Gases Association
- GB/T 8979-2025 Pure nitrogen, high-purity nitrogen and ultra-pure nitrogen (table preview)Standardization Administration of China
- Argon 5.0 product datasheetLinde Gas
- VERISEQ medicinal oxygen: specificationLinde
- Production of oxygen-lean argon from air (patent abstract)U.S. Department of Energy OSTI
Related
- Aviation MRO: oxygen and nitrogen check points (animated)
- Gas-to-liquids and methanol: measurement points (animated)
- Ammonia, urea and air separation: measurement points (animated)
- Industrial gases and compressed air: measurement points (animated)
- Industrial gases: purity checks
- Dew point to ppmV conversion table
- All process maps by industry
Frequently asked questions
Where is moisture measured in an air separation unit?
At the outlet of the front-end adsorbers, to catch breakthrough before water reaches the cold box, and in the products and at cylinder filling as part of the product specification.
How is oxygen purity measured in an ASU?
Percent oxygen in the oxygen product is measured with a paramagnetic or electrochemical analyzer; trace oxygen in nitrogen and argon with a ppm-range electrochemical analyzer.
Is CO₂ at the adsorber outlet a safety issue?
Yes. CO₂ that reaches the cold box can plug the reboiler and let hydrocarbons concentrate in liquid oxygen, so plants alarm at about 1 ppmv.