Aluminium smelters: where to measure moisture, O₂ and CO₂

Follow the air, gas and metal from the compressor house and potline to the gas treatment centre, anode baking furnace and cast house. Each numbered point shows what is measured, the typical value, why it matters and the analyzer that fits.

AIRTO CASTINGFLUE GASPOT GASAIRCOMPRESSORSDRYERSPOTLINE · REDUCTION CELLSGAS TREATMENTCENTREANODE BAKING FURNACEFUME TREATMENTCENTREHOLDING FURNACEIN-LINEDEGASSERAr / N₂N₂ GENERATOR1H₂O2H₂O3CO₂4HF5PFC6O₂7HF8H₂O9O₂10O₂11H12O₂
YORK instrumentIntegrated system (supplied through YORK)

Tap a numbered point. Swipe the diagram sideways on a phone.

Request a quote for this point →
Auto tour
Project list 0 →

⤓ Download the 2-page PDF application note

How the process works

The GCC is home to some of the world's largest primary aluminium smelters. Gas measurement there is less about the electrolysis itself and more about the systems around it: dry compressed air for the potline, the fluoride and carbon balance at the gas treatment centre, combustion in the baking and holding furnaces, and dry, clean degassing gas in the cast house.

  1. Compressed air. Compressors and dryers supply the point feeders, crust breakers and alumina conveying. Dry air keeps moisture out of the cells and the valves working.
  2. Potline. Alumina is reduced in the cells. Hydrogen from moisture in the alumina and air reacts with the bath to form HF.
  3. Gas treatment centre. Pot gas, diluted about a hundredfold with air, is scrubbed with fresh alumina and filtered. The stack is checked for HF, dust, SO₂ and PFCs.
  4. Carbon plant and anode baking. Green anodes are baked in a ring furnace; the pitch volatiles burn as fuel and the fumes are treated before the stack.
  5. Cast house. Metal from the holding furnace passes an in-line degasser, where argon or nitrogen strips hydrogen before casting.
  6. Nitrogen supply. A nitrogen generator or bulk tank feeds degassing, dross cooling and purging.

Measurement points at a glance

#PointMeasuresTypical value / limitRecommended solution
1Compressed air dryer outletPressure dew pointClass chosen by the site under ISO 8573-1, e.g. Class 2 ≤ −40 °C or Class 1 ≤ −70 °CDPT810, DPT830
2Potline air header (point feeders)Pressure dew point at the point of useAs specified by the siteDPT810, DPT-600 Plus
3Gas treatment centre stack / main ductCO₂ (vol %)0.6–1.2 vol % in pot exhaust gasIntegrated system: NDIR CO₂ analyzer with a 0–2 vol % range
4Gas treatment centre stackHF and total fluorideEU BAT-AEL: HF ≤ 1.0 mg/Nm³ (daily average), total F ≤ 1.5 mg/Nm³; GCC permits set their own limitsIntegrated system: in-situ or extractive laser (TDL) HF analyzer
5Pot gas duct / gas treatment centreCF₄ and C₂F₆ (perfluorocarbons)Low-voltage CF₄ about 0.01–1 ppmv; up to 0.5–10 ppmv during anode effects; detection limit needed ≤ 0.01 ppmvIntegrated system: extractive FTIR or quantum-cascade laser analyzer
6Anode baking furnace flue (exhaust ramp)O₂ (%)Set by the furnace operator; high O₂ in the preheat zone signals air ingressPOA3500, DPSS-61XX
7Fume treatment centre outletHF, dust, PAHEU BAT-AEL (integrated plant): HF 0.3–0.5 mg/Nm³ (daily average)Integrated system: laser (TDL) HF analyzer, dust monitor and periodic PAH sampling
8Argon / nitrogen to the degasserDew pointAs dry as possible; no neutral published limit at the point of useDPT830, DPT-600 Plus, DM9600, DM9700
9Argon / nitrogen to the degasserTrace O₂ (ppm)Point-of-use limit set by the cast housePOA200, PGA500
10Holding / melting furnace flueExcess O₂ (%)General guidance for gas-fired furnaces: 2–3 % O₂ (10 % excess air ≈ 2.1 % O₂)POA3500, POA300, PGA530
11Molten metal before castingDissolved hydrogen (mL/100 g)Target < 0.15 mL/100 g in one plant study; porosity thresholds 0.1–0.2 mL/100 gIntegrated system: in-melt hydrogen probe or reduced-pressure test
12Nitrogen generator outletResidual O₂ (ppm or %)Purity set by the use: degassing, dross cooling or purgingPOA200, PGA500, POA300, PGA530

Values are typical figures from published specifications and technical literature. Your contract, licensor or plant specification always governs.

Getting reliable readings

  • Moisture is a hydrogen source. In the potline it feeds HF; in the cast house it feeds porosity. Dry compressed air and dry degassing gas protect both.
  • HF and dust. Raw pot gas carries fluorides and alumina dust. Most raw-gas analyses are integrated laser or FTIR systems; YORK instruments measure the clean utility gases and the conditioned flue gases.
  • Permits. GCC smelters work to national permit limits; the EU BAT values on this page are a reference only.

Frequently asked questions

Why does compressed air dew point matter in a smelter?

Compressed air drives the point feeders, crust breakers and alumina conveying. Moisture it carries reaches the cells, where it is a hydrogen source for HF, and wet air also fails valves and cylinders.

Why must degassing argon or nitrogen be dry?

Water reacts with molten aluminium to release hydrogen, which dissolves in the metal. Wet degassing gas therefore adds the hydrogen it is meant to remove.

Which smelter measurements need integrated systems?

HF, PFCs, dust and SO₂ at the gas treatment and fume treatment centres, and dissolved hydrogen in the melt. YORK supplies them as integrated systems alongside its dew point, O₂ and CO₂ instruments.

Planning analyzers for a smelter?

Send us the measurement points, gas conditions and the parameters you need, and we will propose instruments and sampling systems.